Response Time
t50 < 270 seconds
Repeatability
±10 ppm at 50 ppm
Linearity
±10 ppm from 25-50 ppm, +20% of reading from 50 to 1000 ppm
Temperature Sensor
±0.4°C at 25°C
Humidity Sensor
±3% RH at 50% RH and 25°C
Standard Gases *)
R22, R123, R134a, R404a
Approvals
CE approval, complies with the applicable LVD and EMC directives
(see also Declaration of Conformity)
Certified to UL Std 61010-1 and CSA 61010
Complies with ASHRAE 15-2004