Calibration Procedure

While the ZGARD S & DS sensors are factory-
calibrated, the user must perform regular
calibration checks as part of a routine inspection
and maintenance schedule. Use calibration gases
of known and certified concentrations, and check
the expiration date on the gas cylinders.  

The calibration procedure must be
completed after the replacement of
sensing cells; otherwise the unit could
fail to perform as designed and per-
sons who rely on this product for their
safety could sustain severe personal
injury or death.

Le procès de calibration doit être com-
plété lors du remplacement de cellules
du détecteur.  Sinon, l’appareil peut
faire faillite qui augmente la possibilité
de causer des blessures corporels ou
la mortalité.

Calibration Equipment

Flow controller 0.25 liters/minute and tubing

Purified air as zero gas and Span Gas 
(see TABLE )

SENSOR 

SPAN VALUE

MSA GAS CYLINDER

TYPE

PART. NO.

Carbon Monoxide

60 ppm CO

710882

Nitrogen Dioxide

5 ppm NO2

710332

Combustible

2.5% CH4

10028032

Zero Air

20.8% O2 in Nitrogen

10028042

Calibration magnet (MSA P/N 30060-2)

Calibration cover   (MSA P/N 10128624) 

for S Sensor

Calibration cover   (MSA P/N 10149017) 

for DS Sensor 

Remove the Calibration Cover after
calibration is complete; otherwise, the
unit could fail to perform as designed
and persons who rely on this product
for their safety could sustain severe
personal injury or death.

Enlever le couvercle du calibre seule-
ment une fois le procès de calibration
est terminé.  Autrement, vous courez
chance de faillite de l’appareil et les
gens qui se fient sur cet appareil pour
leurs sécurité cours une chance de
blessures corporelles ou même la 
mortalité.

Calibration Of ZGARD S & DS Sensors 

NOTE: Slow Flash = 1 Flash per second 

Fast Flash  = 8 Flashes per second.

Zero Calibration

1. Apply Zero Air to sensor using 

calibration cover.

2. Apply magnet to calibration target until 

the Green LED slow flashes; 
then, 

remove

magnet.

3. Remove Zero Air when green LED is SOLID

ON (stops flashing)

4. Zero calibration is now complete  

(time required is two minutes);
proceed to Span Calibration.

Span Calibration

1. Apply Span Gas to sensor using 

calibration cover.  

2. Apply magnet to calibration target until 

the Green LED fast flashes; then, 
remove magnet.

NOTE: Green LED slow flashes first. 

3. Remove the Span Gas when the Green LED

is Solid ON. 

4. Span calibration is now complete (time

required is two minutes).

There is a two-minute delay when the
Green LED stays ON.

NOTE: Upon completion of span calibration,

there is a two-minute delay before 
sensor output returns. This time
allows for the span gas to dissipate, 
preventing false indications from 
the controller.

If magnet is left on for more than 16

"

WARNING

"

WARNING

"

ATTENTION!

"

ATTENTION!

4-1

Chapter 4, 
Calibration

Summary of Contents for APOGEE P1

Page 1: ...structions Comme tout appareil complexifié ce produit peut vous donner les résultats anticipés seulement s il est installé usagé et que le service d entretien sont effec tués d après les instructions du fabricant ou manufacturier Sous peine de ne pas suivre les instruc tions ci incluses il est possible que cet appareil vous donne des résultats insuffisant En con séquence les personnes qui dépendro...

Page 2: ...ufacturers of such components THIS WARRANTY IS ON LIEU OF ALL OTHER WARRANTIES EXPRESSED IMPLIED OR STATUTORY AND IS STRICTLY LIMITED TO THE TERMS HEREOF SELLER SPECIFICALLY DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE 2 Exclusive Remedy It is expressly agreed that Purchaser s sole and exclusive remedy for breach of the above warranty for any tortious conduct of...

Page 3: ...SULT ON SERIOUS PERSONAL INJURY OR DEATH 1 Le détecteur d émanations ZGARD S DS tel que représenté dans ce manuel doit être installé opéré et maintenu conformément aux étiquettes les cautions avertissements ainsi que les instructions et les restrictions spécifiées 2 Le détecteur d émanations ZGARD S DS ne doit pas être fixé en plein air ou dans une région où des concentrations explosives de gaz ou...

Page 4: ...e 3 1 ZGARD S DS Sensor Specifications 3 1 ZGARD S Sensors RS 485 Only 3 2 Bump Test 3 2 Table 3 2 Digital Binary Address Codes 3 3 Chapter 4 Calibration 4 1 Calibration Procedure 4 1 WARNING 4 1 Calibration Equipment 4 1 WARNING 4 1 Calibration Of ZGARD S DS Sensors 4 1 Zero Calibration 4 1 Span Calibration 4 1 Sensor Out of Range or Inoperative After Calibration 4 2 Sensor Fail Indication 4 2 Ch...

Page 5: ...emote gas sensors should be strategically placed closest to the areas where the target gases or vapors might occur in the atmosphere 1 1 Chapter 1 General Information and Applications Table 1 1 ZGARD S DS Sensors CARBON CARBON NITROGEN COMBUSTIBLE MONOXIDE SS MONOXIDE EC DIOXIDE EC LEL Principal of Operation MOS Metal Oxide Electrochemical Electrochemical Catalytic Sensor Semiconductor Accuracy at...

Page 6: ...ctric ité du détecteur en entier pour éviter toutes possibilités de chocs qui peut causer des blessures corporelles Do not use the sensor power when connecting any external devices Use shielded cable for wiring installa tion Do not install low voltage signal cable in the same conduit as the other devices such as sensors operating power and or relay wiring Make sure that each sensor is given a uniq...

Page 7: ...ndommager en permanence le détecteur Exécuter tout branchement et installa tion de conduit conformément au National Electric Code A coup sûr si les cautions ci haut mentionnés ne sont pas suivis il y a possibilité de blessures corporelles et vous courez chance d endommager votre propriété 2 2 ...

Page 8: ...SW2 àß 1 TEST FULL SCALE 100 OUTPUT àß 2 TEST HALF SCALE 50 OUTPUT àß 3 TEST ZERO SCALE 0 OUTPUT àß 4 CUT OFF ON OFF SW2 àß 1 TEST FULL SCALE 100 OUTPUT àß 2 TEST HALF SCALE 50 OUTPUT àß 3 TEST ZERO SCALE 0 OUTPUT àß 4 CUT OFF ON OFF ZGARD S Sensors Output Test Each sensor provides a means of driving the output to 0 50 or 100 full scale Respectively when the onboard switch SW2 is tuned ON to 1 or ...

Page 9: ...y Code Setup Each sensor must be given a unique address code for accurate communication The binary address code is set by arranging the user selection switch SW1 as shown Turning on SW1 switches 1 2 3 4 5 6 and 7 according to the appropriate Binary Address Code required for each application See TABLE 3 2 for details 3 2 Bump Test Apply a representative sample gas to the sensor and check if the sen...

Page 10: ...ON OFF ON ON OFF OFF OFF 14 OFF ON ON ON OFF OFF OFF 15 ON ON ON ON OFF OFF OFF 16 OFF OFF OFF OFF ON OFF OFF 17 ON OFF OFF OFF ON OFF OFF 18 OFF ON OFF OFF ON OFF OFF 19 ON ON OFF OFF ON OFF OFF 20 OFF OFF ON OFF ON OFF OFF 21 ON OFF ON OFF ON OFF OFF 22 OFF ON ON OFF ON OFF OFF 23 ON ON ON OFF ON OFF OFF 24 OFF OFF OFF ON ON OFF OFF 25 ON OFF OFF ON ON OFF OFF 26 OFF ON OFF ON ON OFF OFF 27 ON O...

Page 11: ... perform as designed and persons who rely on this product for their safety could sustain severe personal injury or death Enlever le couvercle du calibre seule ment une fois le procès de calibration est terminé Autrement vous courez chance de faillite de l appareil et les gens qui se fient sur cet appareil pour leurs sécurité cours une chance de blessures corporelles ou même la mortalité Calibratio...

Page 12: ...Calibration If Sensor is out of range or inoperative after calibration the Red LED slow flashes Perform a new Zero and Span calibration to resolve the problem Sensor Fail Indication If a sensor fails or is not present the Red LED turns ON and the sensor is not seen by the controller SENSOR FAIL 4 2 ...

Page 13: ...other than authorized MSA service personnel could cause the product to fail to perform as designed and persons who rely on this product for their safety could sustain serious personal injury or death Disconnect all power source s to the ZGARD S DS Sensors before remov ing or changing components Un mot de caution lors de l entretien du détecteur ZGARD S DS Il est nécessaire d utiliser que les pièce...

Page 14: ...A 1 Appendix A Installation Outline Drawings ...

Page 15: ...100 ppm 100mA OUTPUT RS 485 P1 SERIAL 123456 7 Carbon Monoxide Sensor The Safety Company POWER 24VAC Sensor OK Calibration RANGE SENSOR TM ...

Page 16: ...24VAC SERIAL 200mA POWER CO RANGE 100ppm 123456 7 Carbon Monoxide Sensor The Safety Company Sensor OK Calibration NO RANGE 10ppm 2 Calibration Sensor OK Nitrogen Dioxide Sensor SENSOR TM ...

Reviews: