MRU GmbH
SWG 100 User Manual
Page
104
of
141
12.2.
Analyser Status (address 0 and some mirror addresses)
The Analyzer Status is a 32bit-word and must be interpreted bitwise.
Bit
Description
0
Power-On
(until the first zeroing has been done)
1
System-Alarm, see table below
2
Air Purging (zeroing)
3
Gas Sampling (preparing measurement, not measurement!)
4-7
Currently sampled sample point number
(1
10, reads 0 while air
purging)
8-31
reserved for later applications
(read zero)
Some status examples:
Decimal
Hexadecimal
Binary
state description
1
0001h
0000 0001
Power-On (self-test)
5
0005h
0000 0101
First Air Purging (Power-On + Air
Purging)
24
0018h
0001 1000
Preparing
meas.
smp.pt.1
(Gas
Sa smp.pt.1)
16
0010h
0001 0000
Measuring sample point 1
32
0020h
0010 0000
Measuring sample point 2
48
0030h
0011 0000
Measuring sample point 3
18
0012h
0001 0010
Measuring sample point 1 + System-
Alarm
12.3.
Analyser System Alarm (address 2 and some mirror addresses)
The Analyser System Alarm is a 32bit-word and must be interpreted bitwise.
Bit
Description
Meas.
halted
0
Mainboard Offline (some communication problems)
YES
1
Mainboard is in bootloader mode
YES
2
CH4 ambient > threshold value
YES
3
Condensate
YES
4
Sample flow < 20 l/h
-
5
Case fan rotations < 900 rpm
-
6
T-gas cooler > 10°C
-
7
T-gas cooler < 2°C
-
8
T-Sensor > 55°C
-
9
T-Sensor < 5°C
-
10-
31
reserved for later applications
Some system alarm examples:
Decimal
Hexadecimal
Binary
state description
1
0001h
0000 0001
Mainboard is offline, measurement is
halted
8
0008h
0000 1000
Condensate Alarm, measurement is
halted
80
0050h
0101 0000
Sample flow < 20 l/h and T-gas cooler
> 10°C.
Summary of Contents for SWG 100 biogas
Page 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Page 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Page 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Page 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Page 86: ......
Page 92: ......
Page 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Page 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Page 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...