USER MANAUL
MGA
prime Q
MRU GmbH, D-74172 Neckarsulm
27 /
117
5.4.
Device settings
►
Open the menu Settings.
See also Chapter 5.1 Open menu Settings, Page 24.
►
Press Device settings.
A window appears.
In the menu Device settings, you can make the following adjustments
Setting
Range
Description
Analyser start with
This setting must be unlocked first.
For an overview on all available op-
tions please refer to the company’s
home page or sales representatives
.
Source zero point (gas)
Not available for MGAprime-Q
λ
accor. Brettschneider ON / OFF
Calculation of the combustion air ra-
tio (
λ
Lambda) according to
“
Brettschneider
Heated hose
temperature
120
–
180° C
250
–
355° F
Heated probe
temperature
120
–
180° C
250
–
355° F
Ref. temperature
Calculation of the values for a stand-
ard state at reference temperature
Interval auto-zero [h]
See also Chapter Automatic zero
point, S. 54
Modus Slave ID
Modbus address of the instrument
for the remote control via Modbus
Hold delay after purge
Time in which Analog outputs main-
tain their value after zero-point
measurement
Aux connector
Intended for use with the APE acid
dosing unit only.