after you have performed the leak test.
Repeated Zero point
Authorized persons can start a new zeroing after entering the pin code
(Adjustment PIN).
See also chapter 11 Entering Pin code, Page 86.
►
Touch the menu Measure.
►
Touch the Context menu.
A selection list appears.
►
Press Zero point gas.
►
Confirm the zeroing of the gas sensors.
The Zero point gas starts.
The Zero point gas takes 8 minutes
The Zero point gas is taken using the Fresh air inlet port (11).
Automatic zero point
Authorized users can set zero-point intervals after entering the pin code
(Adjustment PIN).
See also chapter 11 Entering Pin code, Page 86.
For measurements over 12 hours, it is recommended to set the zero-
point interval to 12 hours. The zero-point interval can be set in prede-
fined steps.
NOTE
Note that the last zero-point interval set is saved.