18.3.
Switching off the analyser in case error
If the touch screen does not respond, you have two options for switch-
ing off the analyser:
►
Touch the
⏻
button during flashing LED for 30 sec. minimum
►
After change to continuous lighting, release the
⏻
button
The instrument will be switching off with reset.
Alternatively, you can switch off the analyser as follows:
►
Press the reset button on the right side of the analyser.
See also the illustration in 3.4 Connectors, Page 14.
Further information about the defined registers on demand
(
) and on the device supplied USB stick!
18.4.
Connecting analyser with MRU4win
The analyser is a Modbus device.
►
You can connect Modbus devices with MRU4win via a serial interface
(RS 485).
See also Chapter 17.3RS485 Extern, Page 101.
►
You can connect Modbus devices to MRU4win via a TCP/IP
connection.
The TCP/IP connection can be made by LAN or WLAN.
Connecting analyser with MRU4win via serial interface (RS 485).
►
Connect the electrical connections of the RS485 interface as
described in Chapter .17.3RS485 Extern, Page 101.
►
Connect the analyser to your PC via the RS485 interface.
►
Open the Menu Device settings.
See also Chapter 5.4 Device settings, Page 27
►
Set "Modbus Slave ID" to 1.
Please note that the Modbus slave ID is a consecutive number.