14.4 Leak proof test
With the leak proof test the system is checked by the device (incl. the condensate separator) up to
the probe spike on undensity. The internal gas pump generates in addition a subpressure which is
measured over the built-in draft sensor and is observed for a period of 10 seconds. Based on the
decrease of pressure the leakness of the system will be determined.
Operation:
•
Hose set for leak proof set
ATTENTION:
The probe spike before the density test clean! (With depositions on the tube the cap
does not seal.)
•
Launch under “extras” the leak proof test on which the following window will pop up:
OK
…
OR
If of the leak proof test is not passed the probe must be checked including the hosing as well as
the condensate separator.