- 7 -
3.
AT: automatic perform pre-setting data.
(You may use this function when the temp. is
unstable or with big temp. difference)
Operation: 1.
Press SET after 5 seconds and show AL1 or
AL2.
2.
Press SET again and show ∩
「
∪ symbol.
3.
Press ▲up and key in 0001 at SV
4.
Press SET and will start to perform, (The
AT light will be on and means it
’s
working
, the
light will be
off
when it
finishes
.
5.
If you want to stop when it
’s working, you may
key in
0000
at
SV
column when PV column
show
s
∩
「
∪ symbol, or just turn off the
power and restart.
6.
If you are using wider range of
temperature, please record the parameter
value of
“
P
”, “ I ”, “D”…
etc. could
save
your time and key in data directly to use it
again.
□TEMP. □HUMIDITY REVISED VALUE
INSTRUTION
1.
Press SET after 5 seconds and into the first
parameter,
2.
When it shows ALL value then press SET again and will
show ATU until it shows the value of pb.
3.
Use ▲or ▼to adjust. After adjust, press SET again
and wait for 5 seconds, it will return to the original
situation again.
4.
Or just wait for 30 seconds it will return to the
original setting.
Example: Actually temp. 37.1℃, display on 37℃, pb value
add 0.1