5. Product Parameters

Please do not exceed the opening pressure of the safety valve frequently; 
otherwise it will result in damages of the safety valve.

Model

5

329

510

15

19.1

40

0.15

2

0.05

0.09

0.099

0.15

Effective 

capacity

L

Height

MM

Empty Weight

kg

Full weight

kg

Caliber

MM

Static 

evaporation

(L/d)

LN

2

 Output

L/min

Max. working 

pressure 

(MPa)

Standard working 

pressure 

(Mpa)

First stage safety 

valve relief 

pressure 

(Mpa)

Pressure gauge 

indicating range 

(Mpa) 

Second stage 

safety valve 

release pressure

 (Mpa)

O.D

MM

15

404

750

23

35.3

40

0.38

2

0.05

0.09

0.099

0.15

YDZ-5

YDZ-15

YDZ-30

YDZ-50

YDZ-100

K

YDZ-150

YDZ-200

K

YDZ-300

/300K

YDZ-500

/500K

YDZ-240K

30

454

879

32

56.6

40

0.75

3

0.05

0.09

0.099

0.15

50

506

991

54

95

40

1

3

0.05

0.09

0.099

0.15

100

606

1185

75

157

40

1.3

4

0.05

0.09

0.099

0.15

0~0.25

150

706

1188

102

225

40

1.95

6

0.05

0.09

0.099

0.15

200

758

1265

130

296

40

2.4

8

0.05

0.09

0.099

0.15

240

758

1350

148

350

40

2.9

8

0.05

0.09

0.099

0.15

300

857

1459

202

448

40

3.3

8

0.05

0.09

0.099

0.15

500

1008

1576

255

665

40

5.5

10

0.05

0.09

0.099

0.15

The storage area for stainless steel liquid nitrogen biological containers shall have 
sufficient space and shall be well-ventilated.

The temperature in the installation room shall be controlled at the normal temperature 
level of 15°C-30°C.

The relative humidity in the installation room shall be controlled within the range of 
40%-70%.

The indoor floor for placement of stainless steel liquid nitrogen biological containers 
shall be flat.

The positions for placement of stainless steel liquid nitrogen biological containers 
must be far away from the places with heat sources, combustibles, corrosion, intense 
magnetic field, dust and serious oil stains.

The vacuum sealing joints have been sealed before the products are delivered out of the

factory and protected by vacuum sealing joint hoods, and it is strictly prohibited to remove the
vacuum sealing joint hoods privately.

The container is installed with a safety valve. The set pressure of the safety valve has been

adjusted before delivery out of the factory. The user shall not disassemble it in use. If it is
disassembled, the Company has the right and reason not to implement the three-guarantee
policy specified by the state any more.

In order to make the indicated pressure of the pressure gauge accurate, the accuracy shall be

calibrated regularly. Meanwhile, the dial shall not be impacted by any external force, so as to
prevent deformations from causing errors of indicated values.

The pressure increasing valve and the exhaust valve can control the internal pressure of the

container, and the liquid inlet/outlet valve can control the discharge flow of liquid nitrogen. In use,
the handle shall not close the valve excessively, so as to avoid damages; meanwhile, attention
shall be paid to the rotating direction of the valve handle; and the handle shall be rotated
anticlockwise to open the valve and shall be rotated clockwise to close the valve. When the
container is out of use for a long time, the pressure increasing valve must be closed and
meanwhile the exhaust valve must be opened.

When the pressure gauge indicates a pressure reading, it is prohibited to unscrew the lock nut

or the liquid level meter at the top of the valve seat. First the exhaust valve shall be opened for
pressure relief. When no gas is discharged from the exhaust valve and the indicated value of the
pressure gauge is 0Mpa, the lock nut or the liquid level meter shall be unscrewed from the valve
seat;

1. Operating Environment Requirements

2. Precautions for Operation and Use of Products

Environmental 
Requirements

PRODUCT USE

9

8

WARNING

Summary of Contents for YDZ-100K

Page 1: ...leasereadthis instruction carefully before us The company reserves the right to explain the instructions The appearance of the product is subject to the actual product Please keep the invoice together after reading No further notice in case of product technology or software upgrad 2 06 A ...

Page 2: ...uid nitrogen filling port or the level meter to fill the drain port of the liquid nitrogen machine fractionating tower or the pipe 2 The container adopts the boosting method of applying a pressure to the liner after vaporization of a little liquid nitrogen which has the advantages of convenient operation and rapid boosting and guarantees the purity of liquid nitrogen 3 The container has high stren...

Page 3: ...operator After use of the Self pressurized Series tank the inlet valve must be closed The bottom of the container is equipped with four high strength casters which is convenient for short distance movement of the container When moving care should be taken and it must not be pushed or moved violently to prevent collision It is strictly forbidden to move on uneven ground Transfer the Product with ca...

Page 4: ...urface transfer away the sample immediately otherwise the sample can be damaged We bear liability for product loss only and not bear the liability of the third party The container mainly comprises a shell a liner swivel casters drain pipes various valves pressure gauges vacuum sealing joints etc see Fig 1 Fig 2 and Fig 3 For any quality issue please return the Product to the Company for inspection...

Page 5: ...xhaust valve must be opened for venting and pressure relief When the container is out of use this valve shall not be opened Pressure increasing valve When it is opened the pressure of the liner will rise When it is out of use it shall be closed tightly Pressure gauge It is used to indicate the pressure of the liner Liquid inlet outlet valve It shall be connected with the liquid hose to fill liquid...

Page 6: ... vacuum sealing joints have been sealed before the products are delivered out of the factory and protected by vacuum sealing joint hoods and it is strictly prohibited to remove the vacuum sealing joint hoods privately The container is installed with a safety valve The set pressure of the safety valve has been adjusted before delivery out of the factory The user shall not disassemble it in use If i...

Page 7: ... it is required to convey liquid nitrogen all exhaust valves must be open 3 Product Operation and Use ATTENTION After the pressure increasing valve is opened usually the lower part of the shell is frosted which is the normal phenomenon when the container is working When the container is fully filled with liquid nitrogen but a gas phase space equal to 10 of the geometric volume of the container is ...

Page 8: ...e Medium and long distance transport No liquid nitrogen is stored 4 Test and Calculation Method of Daily Evaporation Rate of Product Note The daily evaporation rate of the product shall be tested according to GB T 18443 5 2010 When the ambient temperature and the atmospheric pressure are not tested according to the above regulations or the minimum division value of the measuring instrument is diff...

Page 9: ...ing rod Lower ring connecting rod Measuring rod Power supply 3 6V lithium battery power supply average working current about 15uA Effective liquid level measuring range of sensor 200mm 1000mm Diameter of sensor rod Φ12mm Operating humidity range relative humidity 0 80 Measuring accuracy 1 Working pressure 0 0 6MPa Measuring medium liquid nitrogen Sleeve the PTFE flat seal gasket into the sensor an...

Page 10: ...rd of the current level meter Long press S to display It indicates setting the capacitor value when the current level meter doesn t contact liquid nitrogen Long press S to display It indicates setting the capacitance value when the test range length of the current level meter the upper hole of the level meter is completely soaked in liquid nitrogen Alarm and faults of liquid level detection alarm ...

Page 11: ... ice and can t increase the pressure the following ice melting measures can be taken Adopt the air pressure method to eliminate the residual liquid nitrogen in the container The method is as follows Screw the lock nut at the top of the valve seat tightly close the pressure increasing valve open the liquid inlet outlet valve and the exhaust valve and use nitrogen or dry gas do not use oxygen or any...

Page 12: ...ayer infusion hose Product Instructions PTFE sealing gasket Item Type Item Type Regular Model Qty Qty Unit Unit Remarks Remarks Including locking nuts and pressure stabilizers Standard Standard Standard One piece is equipped and one piece is standby 1 1 1 1 2 K Series Products YDZ 5 15 30 50 are equipped with lock nuts while the regular models YDZ 100K 150 200K 240 300 500 are equipped with float ...

Reviews: