NOVA ST-140
Main Temperature Controller
This control is marked TEMP. CONTROL and consists of the digital display and UP
and DOWN arrow pads for inputting set point temperatures and calibration.
PV-
display of actual temperature inside cabinet.
SV-
Set-up temperature indicator.
AT-
Automatic Calculation P.I.D
OUT-
Control system output indicator
ALM
- Over
-
temperature protection indicator.
Once the indicator is lit, it means the temperature is over the set-up temperature range
for protection.
SET-
If there is a temperature change, push this button first.
Then there is a 4
-
digit number shown on SV, and one of the lamps will be lit. The lamp
can be used to set up the temperature. To push
◄
button to move left, and
▲
button to
increase,
▼
button to decrease.
Operation
1.
Connect piping to the chiller as described on the diagram, depending on using
and type of the appliance.
2.
Fill liquid above the top of the cooling pipes in the bath to avoid damage to both
the heating element and the cooling system.
3.
Make sure that the water level indicator is OFF, if it is ON, fill more liquid.
4.
Connect the main power plug to 230VAC and turn on the unit from the main
power switch.
5.
If the working temperature is 30°C or below 30°C, turn the cooling system on.
6.
Adjust the setpoint temperature, use the UP or DOWN Arrow keys and confirm
with SET button.