NOVA ST-140 

 

 

Main Temperature Controller 

 
 

 

 

This control is marked TEMP. CONTROL and consists of the digital display and UP 
and DOWN arrow pads for inputting set point temperatures and calibration. 
 

PV- 

display of actual temperature inside cabinet. 

SV- 

Set-up temperature indicator. 

AT- 

Automatic Calculation P.I.D 

OUT-

Control system output indicator 

ALM

- Over

-

temperature protection indicator. 

Once the indicator is lit, it means the temperature is over the set-up temperature range 
for protection. 

SET-

If there is a temperature change, push this button first. 

Then there is a 4

-

digit number shown on SV, and one of the lamps will be lit. The lamp 

can be used to set up the temperature. To push 

button to move left, and 

button to 

increase, 

button to decrease. 

 

Operation 

1.

 

Connect piping to the chiller as described on the diagram, depending on using 
and type of the appliance. 

2.

 

Fill liquid above the top of the cooling pipes in the bath to avoid damage to both 
the heating element and the cooling system. 

3.

 

Make sure that the water level indicator is OFF, if it is ON, fill more liquid. 

4.

 

Connect the main power plug to 230VAC and turn on the unit from the main 
power switch. 

5.

 

If the working temperature is 30°C or below 30°C, turn the cooling system on. 

6.

 

Adjust the setpoint temperature, use the UP or DOWN Arrow keys and confirm 
with SET button. 