background image

SI-600 Multi-tube Vortexer                                     

 

              Chapter 4 Operation Guide 

─ 

8

 

 

 

4.  Pulsing mode 

 
 
 
 

 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 

 

a) 

Press “Stop” and “Short” key simultaneously   

into  pulsing  mode,  then  the  sign            disappear 

and red light is on. 

 

b) Press Speed’s or Time’s “

” key and “

” key to set 

the pulsing speed and time. In pulsing mode, the max. 

timer time is 99min59sec. 

 

c) If we set the pulsing time is 20 seconds, then press 

“Start”  key  to  run,  it  will  pulse-on  and pulse off  times 

with 20 seconds in 1 second intervals. 

 

d) Press “stop” key to stop. 

 

e) Press “Stop” and “Short” key simultaneously again 

to  quit  the  pulsing  mode,  then  the  sign          display 

again and red light is off. 

 

 

2400

2400

2400

2400

         

02:00

02:00

02:00

02:00

 

rpm 

m/s 

 

2400

2400

2400

2400

         

0

00

00

00

0::::2

22

20

00

0

 

rpm 

m/s 

2400

2400

2400

2400

         

02:00

02:00

02:00

02:00

 

rpm 

m/s 

Summary of Contents for SI-600

Page 1: ...Operation Manual for Multi Tubbe Vortexer SI 600 PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC VER 01 8 12 ...

Page 2: ... In order to use the instrument properly please read this manual carefully before using the Instrument Opening Check Please check the Instrument and Appendix with the packing list when you first open the instrument packing case If you find there is something wrong with the Instrument and the Appendix do contact the vendor or the producer ...

Page 3: ... style I type GB9706 1 Before using the device read the Manual carefully These units are designed for use in laboratory environments The device must be used by skilled personnel with the appropriate training The operator should not open or repair the Instrument by himself which will result in losing the qualification of repair guarantee or occur accident If there is some wrong with the Instrument ...

Page 4: ...nd contact the vendor There is some liquid flowing into the Instrument Drenched or fire burned Abnormal operation such as abnormal sound or smell Instrument dropping or outer shell damaged The function has obviously changed Before power on guarantee the voltage used should be accordant to the voltage needed and the rated load of electrical outlet should not lower than the demand If the electric li...

Page 5: ...d the function 2 Chapter 3 Preparations 3 1 Structure Description 3 2 Keyboard and Display panel 4 3 Power connection 5 4 Exchangeable Tube Foam Rack 5 Chapter 4 Operation Guide 6 1 Speed and time setting 6 2 Key combination 7 3 Pulsing mode 8 Chapter 5 Failure analysis and handling 9 ANNEX 1 Wiring Diagram for SI 600 10 ...

Page 6: ...The multi tube vortex mixer is a tool to provide that extra degree of mixing however it accommodates full racks of tubes or microwell plates as opposed to singe tubes Consequently this unit is useful for higher throughput mixing of samples Features LCD display it is easy to setup and use Procedure restored function when it power down suddenly and then re power on the device can return to the set p...

Page 7: ...perating condition Ambient temperature 4 C 45 C The relative humidity 70 Power supply AC100 240V 1 5A 2 The basic parameters and function Model Parameter SI 600 Speed 500 2500rpm Orbit 3 6mm Time setting 1min 99h59min Max load 4 5kg Input power 60W Outer dimension mm 426 250 480 Weight kg 14 5 ...

Page 8: ...exer s mechanical structure and the central panel s functions and some preparations before power on You should be familiar with this chapter before the Multi Tube Vortexer is first operated 1 Structure Description Support plate Mains power supply Power connector Screw knobs Foam tube rack Panel USB input interface ...

Page 9: ... the highest speed by pressing this key continuously and will stop running when leaving go of it Start key short pressing this key to start the procedure Stop key short pressing this key to start the procedure Increasing key for speed and time it can setup speed and time Decreasing key for speed and time it can setup speed and time ...

Page 10: ... the machine The external power supply unit power adapter AC100 240V ensures the external safety of the unit 4 Exchangeable Tube Foam Rack Type Parameter Tube Capacity Dimension mm SI B01 10mm 50 245X132X45 SI B02 12mm 50 245X132X45 SI B03 13mm 50 245X132X45 SI B04 16mm 15ml 50 245X132X45 SI B06 25mm 15 245X132X45 SI B07 50ml 15 245X132X45 SI B00 Replacement tray pad set 305X178 5X25 AC100 240V ...

Page 11: ...ase or decrease and also increase or decrease in hundred digits and thousand digits For instance press the key and hold on for 2s the digits will be moved from tens to hundreds increasing the alteration range d Press time s or key units will increase or decrease and also increase or decrease in tens hundreds and thousands For instance press the key and hold on for 2s the digits will be moved from ...

Page 12: ...save automatically The way to switch from m s to h m is the same to above b Key lock function Press speed s key and time s key simultaneously to activate the key lock funtion The key lock prevent from setting time and speed during a mixing operation When open the key lock you can set the parameters again After setting the parameters in display flicks three times and saves the changes automatically...

Page 13: ...r time is 99min59sec c If we set the pulsing time is 20 seconds then press Start key to run it will pulse on and pulse off times with 20 seconds in 1 second intervals d Press stop key to stop e Press Stop and Short key simultaneously again to quit the pulsing mode then the sign display again and red light is off 2400 2400 2400 2400 02 00 02 00 02 00 02 00 rpm m s 2400 2400 2400 2400 0 0 0 00 0 0 0...

Page 14: ...n of power Switch Failure Exchange the switch 1 No display Others Contact the seller 2 Shaking too big Mixing samples are placed asymmetrically Place the mixing samples symmetrically 3 The actual speed is different from the displayed speed Failure in control board Contact the seller 4 ERR in the display Shaking speed is out of control Contact the seller 5 Press invalid Press key failure Contact th...

Page 15: ...600 Multi tube Vortexer Annex 1 10 Annex 1 Wiring Diagram for Multi Tube Vortexer SI 600 M otor L PE N Earth 24V 24V N PE L J2 J3 DC24V MOTER Key and display Key and display PORT1 AC100V AC240V Main Board ...

Reviews: