Operation

 

Manual

PCR-IF-8C

Isothermal

 

Fluorescence

 

PCR

PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

 Hagavish st. Israel 58817  Tel: 972 3 5595252, Fax: 972 3 5594529   

[email protected]

 

MRC.

2

.22

Summary of Contents for PCR-IF-8C

Page 1: ...Operation Manual PCR IF 8C Isothermal Fluorescence PCR PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 2 22...

Page 2: ...I...

Page 3: ...software operation of the instrument simple fault analysis transportation and storage requirements etc This instrument is non IVD instrument It is limited to scientific research and teaching purposes...

Page 4: ...instrument conform to class of GB 4793 1 standard Suggest use this instrument indoors or outdoors place where avoid light Disassemble the instruments without permission is forbidden This will make yo...

Page 5: ...atform where away from water and dust during the outdoors using The instrument should be kept in a cool clean place and avoid directly sunlight strong light to make sure Fluorescence detection can be...

Page 6: ...tion hole and disinfect the instrument with ultraviolet light before transport please Retain the original package for transport using to avoid damage during transportation Warning label Warning Hot su...

Page 7: ...2 Features 2 Chapter 3 Accessories List 4 Chapter 4 System structure and accessories 5 Chapter 5 Operation Guide 8 Chapter 6 Fault analysis and solutions 17 Chapter 7 Maintenance storage and transport...

Page 8: ...of genetically modified food etc Working principle Temperature control block of the instrument heats the sample seat to heat the tube in the seat to the needed temperature The heating lid function ca...

Page 9: ...em 10 150 L Recommended 20 30 L Power supply Power Max AC100 240V 50 60Hz 100VA Dimensions W D H Max 145 305 100mm Weight Max 2 5kg Temperature control Heating lid function Yes temperature is adjustab...

Page 10: ...Features 3 Fusion function Analysis of melting curve and melting peak Result display mode 1 positive negative 2 peak time 3 Tm value Optional software PC software Software upgrade Picture updated U d...

Page 11: ...form is selected according to the order 3 Power adapter pcs 1 4 Operation manual pcs 1 1 each 5 USB flash disk contain software pcs 1 Software included according to order requirements 6 USB cable USB...

Page 12: ...y introduces the structure of the instrument kit installation as well as the preparatory work before starting up Please have a comprehensive understanding about the instrument before the first time us...

Page 13: ...ation Manual Chapter 4 System structure and accessories 6 4 3 Power adapter 4 4 USB cable USB flash disk 4 5 Thermal printer Disclaimer the above picture is for indication only subject to actual deliv...

Page 14: ...ating lid gently it will be closed well when listening click sounds Fig 4 4 Schematic diagram of sample installation Notice The instrument is forbidden to mix centrifugal tubes in one operation to avo...

Page 15: ...ault is Detection Interface Isothermal amplification 5 2 Fast start experiment Name Button Description Test Store and summarize new settings or running program files read section 5 4 Data Store and su...

Page 16: ...on the pop up dialog the name of the original program file will not change As shown in figure Click OK to enter the running interface 1 Temperature setting Click Test Settings Experiment Setting inte...

Page 17: ...he appropriate experimental channel according to the reagent dye wavelength Channel 1 FAM SYBR Green 4 Setting of peak head time parameters The peak time Tt of the amplification curve is related to th...

Page 18: ...e changed The selected channel conditions and peak time settings negative and positive settings can be viewed in real time The real time sample holder temperature progress bar remaining time and total...

Page 19: ...ent curve of the hole Click Null and all curves will be hidden Note The hole color can be set in color read section 5 6 4 Anneal curve It is used to display anneal curves in real time which is divided...

Page 20: ...interface the parameters of peak time and negative and positive can be edited and the corresponding results will be updated Data save as Data files saved format File name and saved time Data manage C...

Page 21: ...etting the interface included Experiment program new manage instrument setting and information system setting function etc As shown in figure Experiment setting function please read section 5 3 Sample...

Page 22: ...he original color of the instrument Return the previous interface It will remind users if not save Language Setting Choose Chinese or English in Language setting interface click to switch it after sel...

Page 23: ...m setting include Battery Lid Temperature Time Screen Update Maintain Maintain It is a professional option for maintenance personnel or instrument developers to maintain and calibrate instruments Plea...

Page 24: ...ct temperature value Sensor failure Contact with distributor 4 Sample holder heating lid temperature alarm or not work Thermistor failure Contact with distributor 5 The cooling fan doesn t work sounds...

Page 25: ...Capture chip read failed Module temperature Module temperature sensor open Module temperature sensor error Module temperature sensor short Module temperature sensor error Heating lid temperature Lid t...

Page 26: ...s any foreign matter to clean users should wipe it slowly by the balloons Turn off the heating lid to prevent the reaction hole from the dust when not use Wipe the sample hole with clean soft cloth wi...

Page 27: ...ects our company will be responsible for replacement Twelve months from the date of delivery provide warranty for faults due to material and manufacturing defects During the warranty period our compan...

Page 28: ...tions and Tags Abbreviations The following Abbreviations are for reference and will appear in this operation manual A Ampere AC Alternating current DC Direct current V Volt Hz Hertz W Watt USB Univers...