2. SPECIFICATIONS

2-1 General Specifications

Circuit

Custom one-chip of microprocessor LSI
circuit.

Display

LCD size : 51 mm x 37 mm
LCD with green backlight ( ON/OFF ).

Display units

Air vilocity : m/s, km/h, FPM, mph, knots 
Air pressure: psi , inch Hg , inch H2O ,  h PA , kPA
mbar, Kg/cm

2

 , mm Hg , meter H2O , Atmosphere.

Measurement

Air vilocity & Dual differential input, data hold,

Function

zero/relative, memory.

Zero adjust

Push button on the front panel.

Sensor

*

Sensor is built inside the housing.

*

Piezoelectric sensor.

*

Used for dry, non-corrosive and 
non-ionic air and gas only.
Liquid is prohibited.

Datalogger

Auto 

 1 sec to 8 Hour 59 Minute 59 sec.

Sampling Time

@ Sampling time can set to 1 second,

Setting range

but memory data  may loss.

Manual  Push the data logger button 

once will save data one time.

@ Set the sampling time to 

0 second.

@ Manual mode, can also select the 

1 to 99 position ( Location ) no.

Data error no.

 0.1 % no.  Of total saved data typically.

Memory Card

SD memory card. 1 GB to 16 GB.

Advanced 

* Set clock time (Year/Month/Date,Hour/Minute/ Second )

setting

* Set sampling time
* Auto power OFF management
* Set beep Sound ON/OFF
* Decimal point of SD card setting
* SD memory card Format
* Air density setting

2

Summary of Contents for PAM-9212SD

Page 1: ... precision measurement Although this meter a complex and delicate instrument its durable structure will allow many years of use if proper operating techniques are developed Please read the following instructions carefully and always keep this manual within easy reach OPERATION MANUAL PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc...

Page 2: ...me and sampling time information 14 5 5 SD Card Data structure 14 6 Saving data from the SD card to the computer EXCEL software 16 7 ADVANCED SETTING 18 7 1 Set clock time Year Month Date Hour Minute Second 18 7 2 Set sampling time Hour Minute Second 19 7 3 Auto power OFF management 19 7 4 Set beeper sound ON OFF 20 7 5 Decimal point of SD card setting 20 7 6 SD memory card Format 21 7 7 Air densi...

Page 3: ...ro value Microprocessor circuit assures maximum possible accuracy provides special functions and features Super large LCD display with contrast adjustment for best viewing angle setup extra software after execute datalogger just take away the SD card from the meter and plug in the SD card into the computer it can download the all the measured value with the time information year month date hour mi...

Page 4: ... sensor Used for dry non corrosive and non ionic air and gas only Liquid is prohibited Datalogger Auto 1 sec to 8 Hour 59 Minute 59 sec Sampling Time Sampling time can set to 1 second Setting range but memory data may loss Manual Push the data logger button once will save data one time Set the sampling time to 0 second Manual mode can also select the 1 to 99 position Location no Data error no 0 1 ...

Page 5: ...e or heavy duty DC 1 5 V battery UM3 AA x 6 PCs or equivalent AC DC power adapter is optional DC 9V adapter input AC DC power adapter is optional Power Current Normal operation w o SD card save data and LCD Backlight is OFF Approx DC 7 mA When SD card save the data and LCD Backlight is OFF Approx DC 25 mA AC DC power adapter is optional If LCD backlight on the power consumption will increase appro...

Page 6: ...Accuracy m s 4 1 to 100 0 m s 0 1 m s 3 a reading Km h 14 7 to 360 0 km h 0 1 Km h or Mile h 9 1 to 223 7 mph 0 1 mph 1 a mph full scale Knot 7 9 to 194 3 knot 0 1 Knot Air density Ft min 81 19685 ft min 1 Ft min 1 200 a 0 1 m s 0 3 km h 0 2 mile h 0 2 knot 20 ft min Note m s meters per second km h kilometers per hour ft min feet per minute knot nautical miles per hour mile h miles per hour intern...

Page 7: ... h PA 200 h PA 1 h PA K PA 20 K PA 0 K PA inch H2O 80 2 inch H2O 0 05 inch H2O Atmosphere 0 2 Atmosphere 0 001 Atmosphere Unit Max range Accuracy mbar 200 mbar 2 F S psi 2 9 psi Kg cm 2 0 2 Kg cm 2 Note mm Hg 150 mm Hg 23 5 inch Hg 5 91 inch Hg F S full scale meter H2O 2 040 meter H2O Included linearity h PA 200 h PA hysteresis and K PA 20 K PA repeatability inch H2O 80 2 inch H2O Atmosphere 0 2 A...

Page 8: ...Remark Measuring Display unit unit psi PSI inch Hg In Hg inch H2O In H2O h PA h PA KPA _ PA mbar bAr Kg cm 2 _ g C2 mm Hg Hg meter H2O t H2O Atmosphere AtP 6 ...

Page 9: ...dapter Input Socket 3 4 REC Button 3 15 Battery Compartment Cover 3 5 UNIT Button Button 3 16 Battery Cover Screws 3 6 Function Button Button 3 17 Stand 3 7 Set Button Time Button 3 18 Tripod Fix Nut 3 8 Log Button Enter Button 3 19 Plug quick coupler 3 9 P1 Input Socket 3 20 Pitot Tube 3 10 P2 Input Socket 3 21 soft tube 3 11 SD card socket 3 22 soft tube 7 ...

Page 10: ... 3 5 Fig 1 to Select the desired Air vilocity units m s Ft min Km h Knot Mile h 4 Zero adjusting pressing and holding the the Button 3 5 Fig 1 and Button 3 6 Fig 1 for at least 1 5 seconds to the display reading will show zero 5 Pitot tube sensing soft tube 3 22 Fig 1 Install to P1 Input Socket 3 9 Fig 1 and sensing soft tube 3 21 Fig 1 Install to P2 Input Socket 3 10 Fig 1 6 The meter is build th...

Page 11: ...djusting pressing and holding the the Button 3 5 Fig 1 and Button 3 6 Fig 1 for at least 1 5 seconds to the display reading will show zero 5 Install the measuring pipe to Plug quick coupler 3 19 Fig 1 6 The meter is build the two input socket P1 input socket P2 input socket for accepting the differential pressure input Connecting the pipe along the Plug 3 19 Fig 1 to a P1 input socket 3 9 Fig 1 on...

Page 12: ...tton 3 4 Fig 1 once to start the Data Record function and there will be a REC symbol on the display 2 With the REC symbol on the display a Press the REC Button 3 4 Fig 1 once the REC MAX symbol along with the maximum value will appear on the display b Press the REC Button 3 4 Fig 1 again the REC MIN symbol along with the minimum value will appear on the display c To exit the memory record function...

Page 13: ...R ZERO ADJUST ON DETERMINE UNIT Optional measuring procedures DATA HOLD MEMORY RECORD RS232 OUTPUT Max Min datalogger function Auto datalogger function or Manual datalogger function Power management AUTO POWER OFF or MANUAL POWER OFF Not activated during Memory Record Selection 11 ...

Page 14: ...ng If the meter is used at first time it should to adjust the clock time exactly please refer chapter 7 1 page 18 d Decimal format setting The numerical data structure of SD card is default used the as the decimal for example 20 6 1000 53 But in certain countries Europe is used the as the decimal point for example 20 6 1000 53 Under such situation it should change the Decimal character at first de...

Page 15: ...n the DATALOGGER symbol will flashing c Finish the Datalogger During execute the Datalogger function press the LOG Button 3 8 Fig 1 1 5 seconds continuously again will finish the Datalogger function the DATALOGGER text will be disappeared and finish the Datalogger function 5 3 Manual Datalogger Set sampling time 0 second a Set sampling time is to 0 second Press the LOG Button 3 8 Fig 1 1 5 second ...

Page 16: ...Datalogger 5 4 To check the time and sampling time information During the normal measurement not execute the Datalogger If press Time 3 7 Fig 1 once the LCD display will present the time information of Year Month Date Hour Minute Second and the Sampling time information 5 5 SD Card Data structure 1 When the SD card is used into the meter the SD card When the first time the SD card is used into the...

Page 17: ... the total files more than 99 files will generate anew route such as PMA02 4 The file s route structure PMA01 PMA01001 XLS PMA01002 XLS PMA01099 XLS PMA02 PMA02001 XLS PMA02002 XLS PMA02099 XLS PMAXX Remark XX Max value is 10 15 ...

Page 18: ...e SD card into the SD card adapter then connect the SD card adapter into the computer 3 Power ON the computer and run the EXCEL software Download the saving data file for example the file name PMA01001 XLS PMA01002 XLS from the SD card to the computer The saving data will present into the EXCEL software screen for example as following EXCEL data screens then user can use those EXCEL data to make t...

Page 19: ...EXCEL graphic screen for example graphic 17 ...

Page 20: ...ON OFF DEC Set SD card Decimal character SD F SD memory card Format dnSt Air density Remark During execute the Advanced Setting function if press HOLD Button 3 3 Fig 1 will exit the Advanced Setting function the LCD will return to normal screen 7 1 Set clock time Year Month Date Hour Minute Second When the upper display show dAtE 1 Use the Button 3 5 Fig 1 or Button 3 6 Fig 1 to adjust the value S...

Page 21: ...e Button 3 5 Fig 1 or Button 3 6 Fig 1 to adjust the value Setting start from Hour value After the desired value is set press the Enter Button 3 8 Fig 1 once will going to next value adjustment for example first setting value is Hour then next to adjust Minute Second value Remark The adjusted value will be flashed 2 After set all the time value Hour Minute Second press the Set Button 3 7 Fig 1 onc...

Page 22: ...ith default 2 After select the upper text to YES or no press the Enter Button 3 8 Fig 1 will save the setting function with default 7 5 Decimal point of SD card setting The numerical data structure of SD card is default used the as the decimal for example 20 6 1000 53 But in certain countries Europe is used the as the decimal point for example 20 6 1000 53 Under such situation it should change the...

Page 23: ...in the Display will show text Ent Sd F to confirm again if make sure to do the SD memory card format then press Enter Button once will format the SD memory clear all the existing data that already saving into the SD card 7 7 Air density When the lower display show dnSt At this time first digit of will flash Use the Button 3 5 Fig 1 or Button 3 6 Fig 1 to adjust the value Air density After the desi...

Page 24: ... instrument become inaccurate 2 Loose the screws of the Battery Cover 3 16 Fig 1 and take away the Battery Cover from the instrument and remove the battery 3 Replace with DC 1 5 V battery UM3 AA Alkaline heavy duty x 6 PCs and reinstate the cover 4 Make sure the battery cover is secured after changing batteries 10 SYSTEM RESET If the meter happen the troubles such as CPU system is hold for example...

Page 25: ...ig 1 The data output is a 16 digit stream which can be utilized for user s specific application A RS232 lead with the following connection will be required to link the instrument with the PC serial port Meter PC 9W D Connector Center Pin Pin 4 3 5 mm jack plug Ground shield Pin 2 2 2 K resister Pin 5 23 ...

Page 26: ...splay mbar 86 psi 23 m s 08 mm Hg 78 inch Hg 80 Km h 10 inch H2O 25 Kg cm2 77 ft min 11 Atmosphere 26 meter H2O 79 mile h 12 K PA 88 h PA 91 Knot 09 D10 Polarity 0 Positive 1 Negative D9 Decimal Point DP position from right to the left 0 No DP 1 1 DP 2 2 DP 3 3 DP D8 to D1 Display reading D1 LSD D8 MSD For example If the display reading is 1234 then D8 to D1 is 00001234 D0 End Word RS232 FORMAT 96...

Page 27: ...U801 WIN is a multi Acquisition displays 1 2 4 6 8 displays software powerful application software SW U801 WIN provides the functions of data logging system text display angular display chart display data recorder high low limit data query text data recorder high low limit data report chart report xxx mdb data file can be retrieved for EXCEL ACESS wide intelligent applications Power adapter AC 110...

Page 28: ...ter SD card structure already get patent or patent pending in following countries Germany Nr 20 2008 016 337 4 JAPAN 3151214 TAIWAN M 358970 M 359043 CHINA ZL 2008 2 0189918 5 ZL 2008 2 0189917 0 USA Patent pending 26 ...

Reviews: