the button . The icon RCF will appear on the screen, and the RCF value is flash on the screen.
Push the button or to set RCF value. Push the button or any other function button to
enter the RCF value.
6. Push the button
, the time value is flash on the screen. Push the button or to set
seconds. Push
again to set minutes. Push the button or any other function button to enter
the time value.
7. Push the button . The rotor type value is flash. Push or to set rotor type (refer to the
below table. ) Push the button again. The tube adapter size value is flash. Push the button
or to choose the tube adapter size. Push the button to enter the rotor value.
Rotor type
Rotor type on the screen
Tube adapter size
AR-1512
2ml, 7ml, 15ml
AR-1524
2ml, 7ml, 15ml
UAR-1512
2ml, 7ml, 15ml
AR-5006
50ml
SB071550
7ml, 15ml, 50ml
8. After setting parameters, push the button
to start the centrifuge. During operation, push the
button to stop the centrifuge.
9. Push the button
to start the centrifuge directly. Release the button and the centrifuge will stop
immediately. The centrifuge will run to the current speed setting value.
10. Brake setting: the initial setting is auto brake and the screen will show
BRK
. If you want to cancel the