background image

MDCEN-302-SMD

 Microprocessor Centrifuge

 

User Manual

PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

 Hagavish st. Israel 58817  Tel: 972 3 5595252, Fax: 972 3 5594529   

[email protected]

 

MRC.

7

.22

Summary of Contents for MDCEN-302-SMD

Page 1: ...MDCEN 302 SMD Microprocessor Centrifuge User Manual PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 7 22...

Page 2: ...unded outlet Turn on the main power switch on the right side of the centrifuge 2 Push the button and then the screen is illuminated Press the button to open the lid The icon is flash on the screen Ins...

Page 3: ...ton again The tube adapter size value is flash Push the button or to choose the tube adapter size Push the button to enter the rotor value Rotor type Rotor type on the screen Tube adapter size AR 1512...

Page 4: ...s no number beside the icon Note Another way to exit the memory mode Push the button to turn off the power and then push the button again to exit the memory mode 12 Under the conditions of power failu...

Page 5: ...ing can be done with a damp cloth Avoid the use of solvents as they may damage the product housing rotor 2 Check if the rotor is in the center In case the rotor is slanted Stop using machine and conta...

Page 6: ...the rotor 3 Contract the distributor for repair Over voltage 1 Rotor type setting is not correct 2 Switching power is defective Turn the main power switch off and then turn it on again 1 Please check...

Page 7: ...s crashed 1 Contract the distributor for repair 2 Contract the distributor for repair 3 Turnoffthemainpowerswitch and turnonthepoweragain Noise is too loud 1 Rotor is unbalanced 2 Rotor type setting i...