Operating Instructions for 

INE-SGW-1 

Automatic Polarimeter

 

PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 

3, Hagavish st. Israel 58817    Tel: 972 3 5595252, Fax: 972 3 5594529     

[email protected]

 

MRC. 

10.

1

9

Summary of Contents for INE-SGW-1

Page 1: ...Operating Instructions for INE SGW 1 Automatic Polarimeter PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 10 19...

Page 2: ...PLES OF APPLICATION 3 3 2 OPTICAL ZERO PRINCIPLE 4 3 3 INSTRUMENT STRUCTURE 6 IV OPERATION INSTRUCTIONS 8 4 1 WORKING CONDITIONS 8 4 2 SETTINGS AND KEYS 8 4 3 COMMUNICATION WITH PC AND SOFTWARE APPLIC...

Page 3: ...urce avoiding the trouble of frequent replacing sodium lights The large screen LCD display on the instrument simple and intuitive stable and reliable II Main Technical Parameters Measuring modes optic...

Page 4: ...ight turned is called optical rotation If the plane polarized light goes through the optically pure substance and the optical rotation is determined by the following three factors a The wavelength of...

Page 5: ...he formula 2 C t t L 2 If the solution contains a non optically active other than the optically substance with concentration of prepared solution and formula 2 the concentration of the optically activ...

Page 6: ...quency and a 2f alternating light intensity b When the quadrature position is deviated from the right the curve C and C in Figure 1 can be obtained The light intensity signal is the superposition of t...

Page 7: ...m of the instrument The light emitted by the light emitting diode passes through the diaphragm condenser lens polarizer a Faraday modulator and collimator in order and forms a beam of light which is v...

Page 8: ...arized light generated by the analyzer reaches a quadrature position the electrical signal with frequency f disappears and the servo motor stops As the instrument starts working the polarizer follows...

Page 9: ...ld be kept in dry environment free from moisture and corrosive gas Make sure to operate the instrument under the environment of 20 3 The instrument uses 220V 50Hz AC power supply AC electronic voltage...

Page 10: ...9 Figure 4 1 operation interface 1 press ReMeas to test the sample See Figure 5 Figure 5 2 press stop to stop test the sample see Figure 6...

Page 11: ...sample see Figure 7 Figure 7 If the measurement is finished yet send will instead of wait key show on display how to use send see 4 3 2 Measuring modes 4 modes optical rotation specific rotation conce...

Page 12: ...d input Length of test tube dm Concentration ReMeas Tims Figure 9 3 concentration press into concentration mode see Figure 10 Need input Length of test tube dm C n If is negative you need not input it...

Page 13: ...12 Figure 10 4 sugar degree Press C it shows as in Figure 11 Figure 11 Press Z back to optical rotation mode See Figure 12...

Page 14: ...specific rotation when specific rotation mode press AV will show the data of optical rotation when concentration mode press CAV will show the data of optical rotation Press these keys again back to t...

Page 15: ...4 4 press Meastimes to change the test time see Figure 15 Figure 15 Notice Select the measurement time between 1 and 6 Select 6 if it is necessary to automatically calculate the mean square deviation...

Page 16: ...nication port USB and RS232 The default is RS232 Press RS232 to change to USB see Figure 16 Press USB to change to RS232 Figure 16 5 Para set 1 Press Para Set see Figure 17 Figure 17 2 press Time see...

Page 17: ...ss otherwise the readings may be affected When placing the test tube care should be taken for the marked position and direction b Take out the test tube then inject the sample to be measured into the...

Page 18: ...te the installation select manual installation that is under the given path find the driver Point the path to the CD ROM provided by the factory Locate Driver folder and select stmcdc inf Follow the c...

Page 19: ...than 1 or greater than 1 the scale factor c Click on save and exit you can save the currently displayed page sheet in the same layout as an individual picture After save the system will exit NOTE Othe...

Page 20: ...corrosive gases and free of severe vibration After using for a period of time due to the influence of the external environment the surface of the optical system may gather dust or mildew which will af...

Page 21: ...factory for repair 2 Replace the source lamp or send back to the factory for repair 3 Replace the 2A fuse No response when pressing Clear 1 The key contact is poor 2 The connection is poor 3 The coun...