When placing the rotor, ensure the full contact between the rotor and the main shaft.

Load the rotor on to the shaft to ensure rotor is in position until it is well connected to the main shaft.

If the rotor is loaded properly on the shaft, you will feel a “click” sound otherwise it is understood that

the rotor is not secured properly on the main shaft. Check once again and if necessary clean it again.

Rotate the rotor slightly with your fingers to check if the rotor vibrates, if so attach the rotor again.

Rotate the nut clockwise using the wrench to tighten the rotor to the shaft firmly.

Rotate the rotor lid in clockwise to the rotor and close  the  rotor  lid and  ensure  t h a t  i t   is  in position.

Close  the centrifuge lid and start operation.

Remove the rotor after operation same as above by turning the locking nut counterclockwise.

8.1.2 Set operating parameters 

Press  the  select  button 

  gently  to  select  the  parameters  and  set  the  parameter  values  to  desired 

setting.  Use  the  parameter  adjustment  knob 

  to  set  the  selected  parameters  (whichever  parameter 

icon flashes). Rotate the parameter adjustment knob in clockwise/anticlockwise direction to increase or 

decrease the parameter values to achieve the desired setting. The minimum speed increment is 10rpm, the 

minimum time increment is 1 second. 

(1)

 

Set the rotation speed

Press the select button

to select the rotation speed in rpm. 

The speed value flashes indicating that the speed is in adjustment mode

The minimum set speed is 500rpm, and the minimum step is 100rpm.

Turn  the  parameter  adjustment  knob

clockwise/anti-clockwise  to  increase/decrease  the 

parameter value. 

(2) Set the running time

Press the select button

to make the time value flash and enter into the time setting mode. 

Turn the parameter adjustment knob

to input the set time within a range of 10sec-99min. 

The centrifuge enters into continuous run mode if the time indicates HD.

Caution

 

Place the rotor on the main shaft and ensure full contact between the rotor and the main shaft. Screw

up  the  nut  on  the  rotor  tightly  using  a  wrench  to  connect  the  rotor  with  the  main  shaft  firmly,

otherwise the rotor might fall off, causing damage to the centrifuge.

Tighten the lid and rotor firmly.

Summary of Contents for HCEN-24

Page 1: ...HCEN 24 Economical Capillary Centrifuge User Manual PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 2 22...

Page 2: ...e the balanced centrifuge tube 4 7 4 Check rotor 4 7 5 Symmetrically load centrifuge tubes into rotor 4 8 Operation 4 8 1 Normal operation 4 8 2 RCF operation 7 8 3 Transient operation 8 9 Cleaning Ma...

Page 3: ...ein are not strictly complied with NOTE Notes indicate an area or subject of special merit emphasizing either the product s capability or common errors in operation or maintenance Do not use this cent...

Page 4: ...s placed on a solid flat and firm table Ensure that the centrifuge is on flat level table before operation Ensure that the angle between the housing and upper lid is larger than the 70 degrees while o...

Page 5: ...the rotor is provided with a lid ensure it is secure it tightly before operation If there is any abnormality observed during operation Ensure to stop the centrifuge immediately contact the service ce...

Page 6: ...functions Rotation speed acceleration switchover transient operating performance Current run time parameter display and sound reminder 2 Scope of application Scope of application suitable for separat...

Page 7: ...20 5 avoid placing the centrifuge in direct sunlight if temperature 3 exceeds 30 4 Ensure to clear 10cm from the centrifuge on both sides and behind the centrifuge in order to ensure air cooling effic...

Page 8: ...Rear view of centrifuge 6 Operating panel Fig 6 1 Schematics of operating panel Power switch Power socket Vent Sight glass Upper cover Speed display Door lock display Time display 1 3 4 2 120 x 100rpm...

Page 9: ...ss this knob to choose the desired speed or acceleration Fig 6 2 Schematics of main display The main display is shown in Fig 6 2 in which the speed is set at 12 000rpm and time is set to 60 minutes Th...

Page 10: ...otor on this centrifuge 7 5 Symmetrically load centrifuge tubes into rotor Caution Ensure to secure the rotor on the main shaft firmly Otherwise the rotor might fall off while the centrifuge is in ope...

Page 11: ...plays the last operating parameters as shown in Fig 8 3 Fig 8 3 Last operation interface Speed setting 12000 rpm Time setting 60minutes The upper lid is closed 8 1 1 Rotor installation and replacement...

Page 12: ...nob to set the selected parameters whichever parameter icon flashes Rotate the parameter adjustment knob in clockwise anticlockwise direction to increase or decrease the parameter values to achieve th...

Page 13: ...o accumulate 3 Error message The centrifuge will automatically stop if any failure occurs in the running process with the failure code indicated on the time display window By looking up table 10 1 the...

Page 14: ...e enters into the operation mode and displays the last set parameter values and allows to reset the target rotation speed 3 Press and hold the pulse key to increase the rotation speed until the set ro...

Page 15: ...dried The rotor needs periodical maintenance It is suggested that the rotor should be cleaned once every three months to ensure cleanliness of the tube hole and main shaft hole before applying a thin...

Page 16: ...cover fails The door opens during operation Press the key if the door is opened Immediately close the outer lid Close the lid before operation E 03 Wrong rotor identification Rotor can not be identifi...

Page 17: ...k button on the right side of the machine At this time the upper cover may be turned open 2 When the centrifuge is turned OFF press the door lock button on the right side of the machine to open the up...

Page 18: ...entrifuge tube Caution Carefully read the user manual and correctly install and use the rotor Don t exceed the maximum speed allowed by the rotor test tube and adaptor The maximum speed allowed by cer...

Page 19: ...er drying 4 It is recommended for regular rotor maintenance and should be cleaned every 3 months to keep the tube holes and shaft clean 5 Apply a thin coat of silicon grease 11 2 Centrifuge tube 11 2...

Page 20: ...r horizontal manner it will deform due to the effect of gravity 2 Remove the threaded cover and inner cover to prevent deformation or crack of the centrifuge tube 3 Take the centrifuge tube only when...

Page 21: ...te of delivery Don t use any rotor damaged due to corrosion or fatigue The damage to the entire unit or rotor due to any of the following reasons is outside the scope of warranty 1 Damage due to impro...

Reviews: