MRC AC-250H Instruction Manual Download Page 3

 

Please   read   this   instruction   manual   before 

using

 

Keep this manual for reference

ACCESSORISE:

1. one piece main machine  2. one piece soundproof lid  . Power supply wire  4. 
one   piece   mesh   basket   (optional)   5.   one   piece   filter   mesh   (   for   outflow 
function). 

APPLICATIONS:

Service   line,   Electron   line,   Medicament   line,   Laboratory,   Machine   line, 
Horniness alloy line, Chemistry line, etc

PREPARATION:

1

Unpack the packing and check if any part get loose. 

2

Keep the cleaner under ventilated and dry condition, place it flatly.

3

Make sure the plug connection is correct and firm. (Use triplex plug and earth 

connection is required and safer.)
4

It’s optional to add certain amount of detergent when needed. 

SPECIFICATIONS:

Model

AC-200H

Power supply

AC220V~2

3

0V 50Hz/60Hz

Frequency

4

0

KHz

Power

240W

Transducer QTY 4(Set)
Tank size

300×240×150

mm(L

×

W

×

H)

Overall size

330×

270

×

310 mm(L

×

W

×

H)

Set time

1-20 minutes

Heating power

500W

Set temperature 

20 -80

Cleaning basket

Have

Lid

Have

Outflow

Have

Setting 

Mechanical control

OPERATION:
1.

START UP ULTRASONIC FUNCTION:

1.

Fill stainless steel tank with water

2.

Plug the cleaner into grounded outlet;

3.

Turn the “Timer” switch left wards to “ON” for continuous working. Or 
turn the switch right to set the 1-20 min working time. During working, 
you   will   hear   the   “   sizzling”   voice   ,that   means   the   cleaner   running 

Summary of Contents for AC-250H

Page 1: ...INSTRUCTION MANUAL FOR Ultrasonic Cleaner AC 250H PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC VER 01 06 10 ...

Page 2: ...NO WATER IN THE TANK THE TANK SHOULD BE MINIMUM 75 FULL WITH LIQUID TO WITHIN 2CM OF THE TOP IF NOT THE HEATING ELEMENT WILL BURN 2 DON T TURN ON THE MACHINE IF NO WATER OR LESS WATER INSIDE 3 DON T PUT THE OBJECT DIRECTLY ON THE BOTTOM OF THE CONTAINER ...

Page 3: ...se triplex plug and earth connection is required and safer 4 It s optional to add certain amount of detergent when needed SPECIFICATIONS Model AC 200H Power supply AC220V 230V 50Hz 60Hz Frequency 40KHz Power 240W Transducer QTY 4 Set Tank size 300 240 150mm L W H Overall size 330 270 310 mm L W H Set time 1 20 minutes Heating power 500W Set temperature 20 80 Cleaning basket Have Lid Have Outflow H...

Page 4: ...ank brim Water and high voltage can cause electrical shock Do not operate the cleaner without proper grounding Do not remove the grounding prong on the line cord plug Do not operate the cleaner dry Attentions a While the machine is working normally ultrasonic and tank syntony gives a well proportioned sound and no shudder on the surface of the water yet there is spray made by the tiny bubbles If t...

Page 5: ...h Clean the dirt in the tank after using i Keep the surface of the machine clean When the machine is broken or the ultrasonic weakens it may be caused by ...

Page 6: ...uid in the tank could vapor in the long time period please keep watching the level of the liquid in the tank d Washing liquid spatters into the shaker results in creepage and short circuit and burn the wafer as well e Electrical abnormity will burn the transistor f High voltage brings damages to the circuit and indicating lights etc Any parts is likely to loose during washing process Do not put fa...

Reviews: