Turning the Unit On

 

Press and hold the Right Key for 3 seconds, until 

buzzer beeps, red LED on, unit is powered on. After 

warm up and self-test sequence, unit enters normal 

reading mode, showing real time gas concentration. 

Turning the Unit Off 

In normal reading mode, press and hold the Right 

Key, unit will display 5 seconds count down, with red 

LED flash and buzzer beeps once per second. After 

the last long flash and beep, LCD displays “Power 

Off”, unit is powered off. 

Charging MP400 

Always fully charge the battery before use. Plug charger 

to the bottom of MP400. Unit LCD will show charge 

status. While charging, the LCD displays a battery icon 

cycle from empty to full. When the battery is fully 

charged, the LCD displays a full battery. 

4

 

Normal Mode 

After warm up finished, unit enters normal reading 

mode, start monitoring gas concentration and the 

reading value will be displayed on the LCD screen. 

 

 

 

 

 

 

 

 

 

 

 

User can check some other values like STEL, TWA, 

PEAK and MIN by pressing Right Key, these screens 

will cycle one by one. 

Config Mode 

In config mode, user can do sensor calibration and 
bump test, change sensor alarm limits and set unit 
parameters.

 

Enter Config Mode 

Press and hold the Left and Right Key together for 3 
seconds, the unit will enter config mode. 

 

The config mode is password protected. 

 

The default password is “0000”. 

 

After all four digits are inputted

 

press Left Key to enter config 

mode. 

 

Config Mode Menu 

 

Sensor: Calibration, Bump, Set Span Value. 

 

Alarm: Set High, Low, STEL, TWA alarm. 

 

Datalog: Set datalog interval, clear datalog. 

 

Unit Setting: LCD contrast. 

 

Wireless: Wireless operation. 

 

Exit: Exit from config mode.   

Exit Config Mode 

Press Right Key to move high-light to “Exit” icon, 

then press Left Key to exit from config mode. 

Change Alarm Limit 

All alarm limits can be customized, including high, 

low, STEL and TWA alarm. 

 

To enter alarm limit setting: 

Config mode 

 Alarm Menu 

 Alarm Type. 

 

To change value: 

        Press Right Key to move high-light field, press 

Left Key to change value, hold Left Key down 

will change value continuously.   

 

To change number direction: 

Press Left and Right together to toggle between   

increase(

) and decrease(

).   

 

To finished setting: 

Press Right Key until bottom-left corner show 

“Done”, press Left Key to exit setting. If value 

has been changed, there will be a message to 

confirm saving the new values. Press “Yes” to 

save, “No” to discard. 

 

7

 