MUNI User’s Guide
20
6.2 Event Log Retrieval and Calibration Certificates
The MUNI automatically stores up to 1000 events consistiing of bump test results, calibration
results, and alarm events that last for more than a few seconds. When the storage is full, the
oldest data get overwritten by the newest event.
•
Event logs can be retrieved either directly from the instrument or via the MuniDock.
•
To download the event log to the PC, select “Download Log” on the bottom panel bar. This
process may take a few minutes because datalogging is always on and many files can be
generated. The event files will appear under the “Datalog” tab on the top of the screen.
Below is a sample screen of event log information listing event type, readings, and other
auxiliary information.
•
To export data to a csv file readable by Excel or other spreadsheet software, move the cursor
over the right data panel, click the right mouse button, and then select either the current
Event file or all the stored files (Whole Datalog).
Bump/Calibration/Alarm results recalled from the MUNI
•
To print a Calibration Certificate, right-click the mouse in the right panel and select Generate
Certificate. Enter any desired information such as operator name and cylinder lot number,
and click Print at the bottom.