PRODUCT OPERATION
5.3.3.3 Bottle sampling
If Bottle Sampling test type is selected, additional functions/parameters will appear in place of the options.
Recommended sample volumes are made on screen specific to sample bottle size. The slider scale should be used to define the
sample volume and a manual flush of the fluid MUST be carried out.
Extreme care should be taken to ensure that there is enough fluid remaining in the sample bottle, this can be done visually
through the clear sample chamber on the Bottle Sampling devices.
Figure 5.46
A manual flush must be performed for the Bottle Sampler test. Selection of this mode disables the auto flush sequence
Ensure that enough fluid remains for the programmed sample volume.
Set sample volumes are suggestions from manufacturer.
NOTE
!
5.3.4 Flush time
Item 24, fig. 5.35.
Sliding your finger left and right the icon will respectively decrease
and increase the time that the unit will hold the flush valve open prior
to a test. Tapping the icon will change the time to the point where it is
tapped. The slider can be used to change the time between 30, 60 and
120s. This can also be changed manually by selecting the icon to the
right of the slider (fig. 5.47).
Figure 5.47
Figure 5.48
Automatic Particle Counters
37