background image

- 10 - 

Optical Fiber: 

 

100BaseFX 

Multi-mode 

Single-mode 

Wavelength 

1300 nm 

1310 nm 

Max. TX 

-10 dBm 

0 dBm 

Min. TX 

-20 dBm 

-5 dBm 

RX Sensitivity 

-32 dBm 

-34 dBm 

Link Budget 

12 dB 

29 dB 

Typical Distance 

5 km

a

 

4 km

b

 

40 km

c

 

Saturation 

-6 dBm 

-3 dBm 

a. 50/125 μm, 800 MHz*km fiber optic cable

 

b. 62.5/125 μm, 500 MHz*km fiber optic cable

 

c. 9/125 μm, 3.5 PS/(nm*km) fiber optic cable

 

 

Power Requirements 

Input Voltage: 

Power Supply Type 

Power Consumption  Fuse Rating 

LV-DC 

20 VDC to 72 VDC 

170 mA @ 20 VDC 

3.15A(T) 2 

HV-AC  85 VAC to 264 VAC 

73 mA @ 85 VAC 

3.15A(T) 2 

HV-DC  88 VDC to 300 VDC 

47 mA @ 88 VDC 

3.15A(T) 2 

 

Connection 

Removable Terminal Block 

Overload Current 
Protection 

1.6 A (protects against two signals shorted 
together) 

Reverse Polarity 
Protection 

Present 

Physical Characteristics 

Housing 

Aluminum, IP30 protection 

Dimensions   
(W x H x D) 

122.5 x 90 x 20 mm (4.82 x 3.54 x 0.79 inch) 

Weight   

Product only: 690 g 
Packaged: 875 g 

Environmental Limits 

Operating 
Temperature 

-40 to 85°C (-40 to 185°F) 

Storage Temperature  -40 to 85°C (-40 to 185°F) 
Operating Humidity  5 to 90% 

Regulatory Approvals 

Safety 

UL 60950-1 

EMI 

FCC Part 15, CISPR (EN55032) class A 

EMS 

EN61000-4-2 Edition 1.2: 2001-04 (Level 4) 
EN61000-4-3: 1995+A1: 2001   
IEC 61000-4-3: 2002+A1: 2002 (Level 3) 
EN61000-4-4: 2004 (Level 4) 
EN61000-4-5: 2001-04 (Level 3) 
EN61000-4-6: 2004-11 (Level 3) 
EN61000-4-8: 2001-03 (Level 5) 
EN61000-4-11: 2004-03 (Criteria B) 

Power Automation 

IEC 61850-3, IEEE 1613 

Rail Traffic 

EN50155/EN50121-4 

Warranty 

Warranty Period 

5 years 

Details 

See www.moxa.com/warranty 

 

Summary of Contents for PTC-101 Series

Page 1: ...upport Moxa Americas Toll free 1 888 669 2872 Tel 1 714 528 6777 Fax 1 714 528 6778 Moxa China Shanghai office Toll free 800 820 5036 Tel 86 21 5258 9955 Fax 86 21 5258 5505 Moxa Europe Tel 49 89 3 70...

Page 2: ...101 perfect for ensuring that your Ethernet equipment can withstand critical industrial applications such as in hazardous locations and complies with FCC UL and CE standards Package Checklist The PTC...

Page 3: ...nal block for power input 3 Heat dissipation vents and relay output 4 DIP switch 5 Power input PWR LED 6 Fiber Link Active LED 7 100BaseFX Port ST SC LC connector 8 10 100BaseT X 9 TP port 10 Mbps LED...

Page 4: ...ions for the PTC 101 S SC other models available by request Wiring Requirements ATTENTION Safety First Be sure to disconnect the power cord before installing and or wiring your Moxa Industrial Media C...

Page 5: ...eparate The rule of thumb is that wiring that shares similar electrical characteristics can be bundled together Keep input wiring and output wiring separated We strongly advise labeling wiring to all...

Page 6: ...erminal block receptor which is located on the PTC 101 LV s top panel ATTENTION Before connecting the PTC 101 LV to the DC power inputs make sure the DC power source voltage is stable Communication Co...

Page 7: ...sed to transmit data from device I to device II and the other optical line is used transmit data from device II to device I for full duplex transmission All you need to remember is to connect the Tx t...

Page 8: ...tion is disabled ON Forces the Ethernet port to use Full Duplex OFF Forces the Ethernet port to use Half Duplex 4 Link Fault Pass Through Enable Disable ON Enables Link Fault Pass Through The link sta...

Page 9: ...port 10 Mbps link is inactive 100M Green On Ethernet port 100 Mbps is active Blinking Data is being transmitted at 100 Mbps Off Ethernet port 100 Mbps link is inactive Specifications Technology Standa...

Page 10: ...A protects against two signals shorted together Reverse Polarity Protection Present Physical Characteristics Housing Aluminum IP30 protection Dimensions W x H x D 122 5 x 90 x 20 mm 4 82 x 3 54 x 0 7...

Reviews: