Dial Torque Analyzer Operation & Calibration Instructions

Rev 1.0

Calibration Procedures

1. The accuracy of the dial torque analyzer should only be verified using an appropriate Calibration Beam and Weight Set 

traceable to national standards.

2.  Mount the Dial Torque Analyzer on a suitable, level bench surface.
3.  Load the analyzer using the Beam and Weight Set to give a torque equivalent to 20% of the full scale range and 

compare with indication on the analyzer dial. 

Do not use the memory pointer during verification.

4.  Check the Dial Analyzer at all major graduations and adjust, if need, as described below until all values are /- 2% 

of reading or 1 graduation, whichever is greater.

Adjustments for DM model

The DM model is a delicate instrument with a low torque range, 
therefore please contact the Mountz Service & Repair Department for 
calibration adjustments.

Adjustments for TCA model

If the TCA model doesn’t indicate the true torque applied within the stated
tolerance, then calibration adjustments are required and performed as
described.
1.  Remove back plate from torque analyzer.
2. Perform the required calibration adjustment as follows:

i.

To perform scale adjustment, adjust the main needle position by 
loosening the locking screw (A) and turning the adjusting screw (B). 
Turn adjusting screw CW to raise reading and CCW to lower reading.
Check the analyzer at full scale.

ii.

To perform linearity adjustment, loosen locking nuts (C) and rotate both Eccentric Spring Pivots (D) in the same direction. 
Moving the Spring (E) towards the main spindle will result in higher indicated torque value and vice versa. Once 
acceptable linearity is achieved, perform scale adjustment as needed.

3. Re-adjust the zero point as necessary by adjusting the lower stop set screw (F) and lock screw (G) that rests against the 

pivot arm so that the indicating needle is positioned in the center of the black area on the dial face.

4.  Perform verification procedure to verify adjustments and repeat as necessary.
5.  Fasten back plate to analyzer.