RTSX Torque Sensor Operation Instructions

Rev 2.0 (5/16/2011)

Calibration Procedures
1.  Attach the RTSX securely to a stand that 

keeps the Rotary torque sensor in a stable
position for the calibration process with the 
cable connector toward the top.

Mountz offers Rotary Fixture Stands for
calibrating Rotary Torque Sensors.

Model: RFS-200  (Torque Range 1- 200 lbf.in)
Item # 060234

Model: RFS-500  (Torque Range 5- 500 lbf.ft)
Item # 062032

2. Connect the RTSX to a torque tester/

display. Review the torque range of the 
torque sensor and select the appropriate 
measurement units.

3.  Determine type of calibration to be performed.

Calibration at 3 Pts.

Test at 10%, 50% and 100 of Full Scale.

Calibration at 6 Pts.

Test at 10%, 20%, 40%, 60% 80% and 100 of Full Scale. 

Direction  

Clockwise and/or Counter Clockwise

4.   Select the appropriate Calibration Arm or Wheel. Attach it to the output drive of the RTSX.
5.   Gently connect the Hanger to the Calibration Arm or wheel. 
6.   Load 3 times to minimum 80% FS in direction of operation and reset to zero after loading.
7.   Apply series of increasing torques in direction of operation starting from the lowest test point.
8.   Record readings from the test device at each test point prior to performing any adjustments.
9.   Repeat steps 6-8 in the opposite direction (if required).
10. Perform calibration adjustments. Repeat test as described above until readings at all test points are within tolerances. 
11. Repeat test as described above and record 5 readings from test device at each test point. Compile all necessary details 

to generate test report.

12. Remove old calibration label and place new label on torque sensor. 

Maintenance Procedures
1.  Recommend cleaning cycle is approximately 1,000,000 Revolutions.
2.  Loosen the 4 fixing screw and remove the cover plate.
3.  Use a soft linen cloth, a fine hair brush or oil-free compressed air to clean the 

dust from the slip-rings and the spaces between them. Carefully clean the 
brushes and the plastic part with the spring using a hair brush or oil-free 
compressed air. Also clean the connector.

4.  Measure the brush thickness, it should be more that 0.5 mm. New brushes can 

only be fitted at the manufacturer.

5.  Replace the cover plate carefully and re-tighten the fixing screws.
6.  The Rotary Torque Sensor should be recalibrated every year.

Mountz Service Centers
Mountz, Inc.

Mountz, Inc.

19051 Underwood Rd

1080 N. 11th St

Foley, AL 36535

San Jose, CA 95112

Phone: (251) 943-4125

Phone: (408) 292-2214

Fax: (251) 943-4979

Fax: (251) 292-2733

Wheel Attachment

In/Out Adapter
(per Application)

Pin Adapter (Match Rotary

Transducer Input Feature)

Front Plate

Top knob

Mounting holes or clamp to bench

Side knobs

In/Out Adapter
(per Application)

Pin Adapter (Match
Rotary Transducer
Output Feature)

Rotary Fixture Stand

Contact 

Mountz Inc.

Phone: 408.292.2214

Fax: 408.292.2733

[email protected]

www.mountztorque.com

Looking for fasteners?

www.mrmetric.com

Reviews: