B.  Interpupillary Adjustment. 

 

1.  Whilst looking through both eyepieces (1), move eyepiece tubes (2) on their axis, holding head 

by the eyepiece tubes housing the prisms. (3). 

 

2.  When a full field of view is obtained, and converted into a single image, interpupillary distance is 

correct for your eyes.   

 

3.  Each user must adjust interpupillary distance to match their eyesight 

 

C.  Focussing 

 

1. 

Turn objective turret (5) so that the lowest magnification number 2X is facing the front of the 
microscope, lined up with the eyepiece, and the 4X magnification is located at the two sides, 
perpendicular to the eyepiece. The turret permits a 1/4 turn. Make sure that turret is “clicked” 
into its correct position. 

 

NB:

 Fot ST-37 model, omit steps 2,3 and 4, passing directly to 5. 

 

 

2. 

Place a flat object or a slide on center of stage (9). 

 
3. 

Turn focus knobs (13) to mid range.  

 
4. 

The head holder (11) is mounted on a column (15), and can be moved up or down to 
accommodate the size of the sample.  

 
a. 

Support head (4) with one hand without touching any lens. 

 
b. 

Without letting go of the head, loosen the screw supporting the head holder (12). 

 
c. 

While looking through the eyepiece (1), move the head holder up or down until the object 
appears in focus.  

d. 

Tighten the head holder support screw. 

 
e. 

These steps are not necessary every time the stereomicroscope is used, only when the sample 
is not in the range of focus.  

 
5. 

Turn focus knobs (13) until the image is sharp.  

 

D.  Adjusting the diopter. 

 

On the left eyepiece tube there is a diopter adjustment collar. Its normal position is when the lower 
part of the collar is aligned with the sign marked on the eyepiece tube. 
 
In the case of differences in eyesight: 

 

With the right eye, look through the right eyepiece (1) and adjust focus.  

 
Now use the left eye to look through the left eyepiece, and adjust the focus by turning the diopter 
adjuster, on the left hand eyepiece tube, until a detailed image is achieved. Do not adjust the focus 
with the focus knobs (13). 

 

E.  Changing magnification. 

 

Turn objective turret (5) so that the highest magnification number 4X is facing the front of the 
microscope, lined up with the eyepiece, and the 2X magnification at the two sides, perpendicular with 
the eyepiece. The turret permits a 1/4 turn. Make sure that turret is correctly “clicked” into correct 
position. 

 

 

 

-5- 

Summary of Contents for ST-30 Series

Page 1: ...Instruction Manual ST 30 Series Copyright 05 03 Motic Microscopes European Division...

Page 2: ...d holder support screw 18 Base 1 Eyepiece 10 Diopter adjuster 13 Focus knob 16 Incident illumination switch I 17 Transmitted illumination switch T 4 Head 2 Eyepiece tube 8 Stage sample clips 15 Column...

Page 3: ...e store it for long periods of time or should it become necessary to transport it to a technical service for any repair or maintenance procedure The box should contain the following components All the...

Page 4: ...observation of microscopic specimens slides or thin or transparent samples such as leaves or insect wings The black and white stage plate is used to study opaque objects or for dissection The best con...

Page 5: ...head loosen the screw supporting the head holder 12 c While looking through the eyepiece 1 move the head holder up or down until the object appears in focus d Tighten the head holder support screw e...

Page 6: ...ntenance WARNING FOR YOUR OWN SAFETY SWITCH OFF AND DISCONNECT THE MICROSCOPE FROM ANY ELECTRICAL SOURCE BEFORE ATTEMPTING ANY MAINTENANCE PROCEDURE TO AVOID THE RISK OF ELECTROCUTION CONSULT YOUR DIS...

Page 7: ...bulb has been touched with the hands clean it as this could affect light transmission or the duration of the bulb e Replace the bulb cover and tighten the screw For LED models a Unscrew the bulb prot...

Page 8: ...s as shown in Fig 1 C Mechanical Maintenance 1 Adjusting the tension of the focus Tension comes pre adjusted by the factory The best point of tension is that which permits the focus knobs to move as l...

Page 9: ...se correctly Have repaired by a qualified specialised technician Fuse blown quickly Wrong fuse Replace with appropriate fuse Fuse blown instantly Short circuit Have repaired by a qualified specialised...

Page 10: ...etails of the required revision Warrantee All MOTIC microscopes are warranted against any manufacturing defect for a 5 year period Damage occurring by any unauthorised repair work or occurring through...