MOS MicroSet MS pH 97 Manual Download Page 15

15 

 

10. SAFETY WARNING 

 

 
 

 

1.  Read Manual carefully and understand instructions & directions provided in this manual. 

 

2. Installation, connections, commissioning and service shall be carried out by only qualified and 

authorized person. 
 

 

3.  To protect instrument from any external hazards, customer should take necessary care while 

preparing site  ready before installation. 

 

4. Ensure proper supply voltage (230 VAC) with proper polarity to the instruments, before Powering 

ON instrument. 
 

 

 TROUBLE SHOOTING PROCEDURE

 

 
 

 

SYMPTOMS

 

CAUSE OF FAILURE

 

ACTION TO BE TAKEN

 

No display indication

 

1.Absence of 230 V AC at 

terminal block.

 

2. Loose connection on 

terminals.

 

1.Check 230 V AC power 

supply & rectify the fault.

 

2.Tight the terminal 

connections.

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 

 

Summary of Contents for MicroSet MS pH 97

Page 1: ...MANUAL MS pH 97 pH INDICATING CONTROLLER TRANSMITTER...

Page 2: ...2...

Page 3: ...3 TECHNICAL SPECIFICATION 06 4 TERMINAL CONNECTIONS 07 5 KEY DETAILS 07 6 CALIBRATION PROCEDURE 08 7 TEMPERATURE COMPENSATION 10 8 SLOPE 11 9 SETPOINT 11 10 MODBUS RS485 optional 14 11 SAFETY WARNING...

Page 4: ...the back plate of the instrument Warranty Terms The terms of this instrument that are warranted are described in the techno commercial offer We will make any repairs that may become necessary during...

Page 5: ...er may only be installed commissioned operated and maintained by properly trained and authorized personnel Look at the ordering detail to ensure that the device is delivered according to your order Ch...

Page 6: ...2 RTD PT 100 Temp Sensor through Terminals 3 4 Supply 230V AC 50 Hz Output 4 20 mA DC MODBUS RS485 Optional Temperature Compensation Manual Auto through PT 100 Range 0 149 9 O C Accuracy 0 01 pH Cali...

Page 7: ...Sensor to BNC Connector on back plate 3 Connect RTD PT 100 Temperature Sensor if required to Terminals 3 4 on back plate Please connect RTD PT100 Temperature Sensor AFTER Calibration Procedure 4 Conne...

Page 8: ...starting Calibration Procedure make sure Temperature is in Manual Mode at 25 C When RTD PT100 Temperature Sensor is NOT connected Temperature is in Manual Mode Remove Protective cum storage cap of pH...

Page 9: ...9...

Page 10: ...troller Transmitter MS pH 97 automatically goes in Automatic Temperature Compensation Mode When RTD PT 100 Temperature Sensor is Not Connected pH Indicating Controller Transmitter MS pH 97 is in Manua...

Page 11: ...4 00 Then in this case Relay will get ON when pH value goes below 3 00 pH Relay will get OFF when pH value goes above 4 00 pH 2 You can use SPL2 SPH2 for High range controlling of pH If you enter SPL2...

Page 12: ...12 Similarly for setting of Setpoint SPH2 SPL2 SPH2 please refer below flow chart...

Page 13: ...13...

Page 14: ...Indicating Controller Transmitter MODBUS communication is based on RS485 RTU User can set the salve ID as per the requirement with following Specification Baud Rate 9600 Slave ID 1 default Start Bit...

Page 15: ...external hazards customer should take necessary care while preparing site ready before installation 4 Ensure proper supply voltage 230 VAC with proper polarity to the instruments before Powering ON in...

Page 16: ...16...

Page 17: ...roSet Model MS pH 97 Serial No __ __ __ __ __ Input pH Simulator pH Sensor Model MS pH ___ SN __ __ __ __ READING Standard Buffer Solution pH Observed Reading Before Calibration pH Observed Reading Af...

Page 18: ...18...

Page 19: ...e company reserves the right to decide the workplace for the repair work The freight for defective material will have to be borne by the buyer the transit risk for such material will rest with the buy...

Reviews: