AMBULO
™
2400
BP MONITORING SYSTEM
Ref: 9503-182-01-ENG Rev B1
•
Disconnect the USB cable and begin
patient set-up. The Ambulo 2400
will
restart into PAUSE mode and should
display the correct time.
•
Press and hold the START/STOP button
for five seconds to change the device mode
from PAUSE to AUTOMATIC. Once in
AUTOMATIC mode, the device may display
a countdown in addition to the time.
Patient Set-up
•
Select an appropriate-sized cuff, depending
upon the size of the patient’s non-dominant
arm. Each cuff is labeled with range markers to indicate the minimum and maximum
circumference that can be accommodated by the particular cuff.
•
Place the cuff on the patient’s non-dominant arm such that the artery indicator rests
on the brachial artery. Do not tighten the cuff too firmly; allow a finger’s worth of slack
between the cuff and the arm.
•
The Ambulo 2400 can be worn in four distinct positions, depending upon patient
comfort and anticipated activity:
Obtaining Measurements and Reporting
•
Connect the Ambulo 2400 to your PC using the USB cable included with the ABPM
system. Run the software.
•
From the Main menu, click the ‘Download Data from Device’ icon.
•
Access the available views as needed: Table, Graphical, Actigraphy, Histogram,
Statistical, Comparison and Report.
•
Enter free text notes (pertaining to the patient’s condition, a particular episode, diagnosis,
etc.) in order to make them part of the patient’s record. To enter notes for a patient, select
the “Browse Patients” icon, select the appropriate patient and click the ‘Edit’ button.
•
Print an ABPM data report by selecting the ‘Report View’ function. The format of the
report can be determined in the Software Configuration menu.
www.mortara.com
ISO 13485 CERTIFIED
Ambulo™ is a trademark of Mortara Instrument, Inc.
Microsoft
®
and Windows
®
are registered trademarks
of Microsoft Corp.
9503-182-01-ENG Rev B1
Attention, consult accompanying documents.