Q-Stress 4.5 System Installation Instructions
4
Place the laser printer or chart
recorder on top of the cart.
Note:
Follow printer manufacturer’s
instructions to set up the printer.
5
Plug the four component AC power
cables into the transformer.
(See
for
appropriate plug placement.)
6
While holding the four AC cables,
place the transformer on the
bottom of the cart.
7
Connect an AC cable to the
transformer.
Connect the third standard length
AC cable to the computer tower.
Connect the longest AC cable to the
pre-amp.
Route the two of the standard
length AC cables to the top of the
cart for the monitor and chart
recorder.
Note:
Make sure the preamp is 3
feet or farther away from the
isolation transformer. Electro-
magnetic interference could cause
noisy ECG signals.
Step