Signo T500 Root Canal Length Measurement Function
Operation Instructions
Thinking ahead. Focused on life.
Page 1: ...SignoT500 Root Canal Length Measurement Function Operation Instructions Thinking ahead Focused on life...
Page 2: ...ing the equipment and pay close attention to warnings and notes Keep this manual in a readily accessible place for quick and easy reference Trademarks and Registered Trademarks The names of companies...
Page 3: ...g Between Endodontic Mode and EMR Interlock Endodontic Mode 16 4 4 3 Basic Operations 20 4 4 4 Memory Selection 21 4 4 5 Setting Rotational Motions 22 4 4 6 Setting the Number of Rotations 25 4 4 7 Ca...
Page 4: ...ol 43 5 2 Micromotor 43 5 3 Ultrasonic Scaler 43 5 4 Transport and Storage 43 6 Maintenance and Inspection 44 7 Troubleshooting 45 7 1 Items to be Checked before Requesting Repair 45 7 2 Error Display...
Page 5: ...traindications Prohibitions and Precautions for Use Safety and Hazard Prevention of Electric Medical Devices then operate the equipment with the utmost caution to prevent either damaging the equipment...
Page 6: ...floods lightning etc J MORITA TOKYO MFG CORP will stock replacement parts for the product for a period of 10 years after manufacture of the product has been discontinued We will supply these parts and...
Page 7: ...cause the pacemaker or the Implantable Cardioverter Defibrillator ICD to function abnormally The electro magnetic waves emitted by inverters used for film viewers neon lights etc can interfere with th...
Page 8: ...6 Accessories File Holder 3 Parts Identification and Accessories Quantity Probe Cord 1 File Holder 3 Contrary Electrode 5 Tester 1 Contrary Electrode Probe Cord...
Page 9: ...g Reference Numbers p 10 Canal Length Meter Setting of the position of the Flash Bar p 10 Selecting the Beep Alarm Sound Volume p 10 Flash Bar yellow The apex locator is being selected The display swi...
Page 10: ...ectrode are properly connected to the probe cord That the probe cord is properly plugged into the jack on the shoulder of the chair That all meter indicator bars on the display light up by touching th...
Page 11: ...heprobecordjackontheshoulderofthechair Check that the meter on the display indicates within three bars above or below 1 The meter may jump when the tester is inserted Wait for about one second until t...
Page 12: ...ot canal as a reference point to determine the working length 2 REF 1 USER A Specify the position of the Flash Bar using or The position can be set within a range from Apex to 2 The 0 5 reading on the...
Page 13: ...an accurate measurement cannot be made because of the canal shape unusual cases or poor performance of the equipment Always check the measurement with an X ray Stop using the instrument immediately if...
Page 14: ...h them with water 2 Clip on the hand file with the file holder 1 Press down the file holder button with your thumb 2 Clip the file holder on the metal shaft of the file in the root canal 3 Release the...
Page 15: ...reading on the meter indicates a distance of 0 5 to 1 0 mm from the anatomical apex Use this as an estimate to determine the working length which will di er somewhat depending on the individual tooth...
Page 16: ...ockwise rotation torque rotation counterclockwise rotation Clockwise rotation torque rotation OGP OTR counterclockwise rotation Torque setting p 24 Torque limit Torque limit Trigger torque Meter displ...
Page 17: ...1 Operation mode Clockwise rotation Clockwise rotation Clockwise rotation Counterclockwise rotation Clockwise rotation OGP OTR Counterclockwise rotation Number of rotations 300 300 500 200 1 000 300 3...
Page 18: ...eferentially selected The micromotor selection screen normal cutting is displayed UL H Endo USER A 1 1 100 rpm OFF OFF Endo 1 E Endo 1 0Ncm USER A E UL H The micromotor not compatible with root canal...
Page 19: ...ith root canal length measurement Option Explanation of the micromotor key EMR interlock endodontic mode 5 Press the 10R 1 key to change to the EMR interlock endodontic mode screen Press the 1 1 key t...
Page 20: ...nts for EMR interlock endodontic mode p 14 Canal Length Meter The meter is indicated on the unit that have a root canal measurement function option The file tip position is indicated during root canal...
Page 21: ...indicates where the apical action is being operated and where the file tip of the handpiece is positioned Switching between the variable speed constant speed p 27 To the next page Back to previous pag...
Page 22: ...rotation If the micromotor does not rotate even when stepping on the pedal the micromoter may generate heat and cause burns or low temperature burns Stop stepping on the pedal Pick up the micromotor...
Page 23: ...timing and AI catch option timing can be set in the chair unit settings You can set them in the user s personal settings for the chair unit Foot control operation Switching Memory Option Press the End...
Page 24: ...from penetrating into the file or being broken Counterclockwise rotation TR OGP This motion is a combination of clockwise rotation and counterclockwise rotation and works e ectively for creating a gl...
Page 25: ...fely and e ciently without a ecting the root form to reduce the occurrence of file penetration brakeage ledge or perforation In normal operation the motor continues in clockwise rotation and the opera...
Page 26: ...e limit value When rotational motion is set to OTR Torque value when operating the OTR function Trigger torque value The above torque values can be set only when the rotational motion is set to torque...
Page 27: ...e rotational motion is set to clockwise rotation CW torque reverse TR or counterclockwise rotation CCW Increase or decrease the value using or Enter the value and press the Enter key to confirm If the...
Page 28: ...de Page 1 Change the variable speed ratio to go to the screen to enter the number of rotations 10R 1 interlocking with the root canal length measurement function is indicated if the root canal length...
Page 29: ...tations is the maximum number of rotations EMR interlock endodontic mode Page 1 Switching is not possible during micromotor operation If the rotational motion is set to torque reverse only constant sp...
Page 30: ...lternates between clockwise rotation and counterclockwise rotation Perform calibration after a speed down contra angle handpiece for endodontic treatment has been attached to the micromotor When perfo...
Page 31: ...se the cutting part to heat up Endodontic mode EMR interlock endodontic mode Page 1 Page 1 Switching the Water Flow Timing Water flow can be set for each memory Confirm that the water flow is OFF befo...
Page 32: ...valves on the underside of the table Turn the valve in the direction of the arrow to increase the water flow volume and air flow volume Three way syringe Fourth circuit Fifth circuit Third circuit Fir...
Page 33: ...ment by automatically controlling its rotation at a self set position In endodontic mode apical actions are not available 1 1 100 rpm OFF OFF Endo 1 E Endo 1 0Ncm USER A 10R 1 100 rpm OFF OFF Endo 1 E...
Page 34: ...prevents the micromotor from automatically stopping or rotating counterclockwise so that the operation may continue even after the file reaches the Flash Bar position These functions are available on...
Page 35: ...p position in the root canal 4 4 10 Setting BeepAlarm Sound Volumes Page 2 Endodontic mode OFF Beep alarm sound OFF Beep alarm sound volume 1 small Beep alarm sound volume 2 large Page 2 EMR interlock...
Page 36: ...rightness and H high brightness For devices without a light this function is not available light remains o 4 4 11 Setting the Handpiece Light Page 1 Endodontic mode Page 1 EMR interlock endodontic mod...
Page 37: ...ic Scaler Solfy Option SCALING SOFT AS USER A Page 1 2 USER A Page 2 Switching water flow ON OFF p 38 Setting the handpiece light p 39 Canal Length Meter The meter is indicated on the unit that have a...
Page 38: ...nic scaler selection screen model with built in apex locator 2 The ultrasonic scaler selection screen is displayed 3 Step on the foot control pedal to start the micromotor and release the pedal to sto...
Page 39: ...sh Bar position SwitchingtheApicalStopFunction In this setting you can switch vibration ON OFF when the file tip reaches a self set position in the root canal the Flash Bar position Apical stop ON Thi...
Page 40: ...Foot control operation Switching water flow ON OFF Adjusting Water Flow Adjust the water flow volume using the valves on the underside of the table Turn the valve in the direction of the arrow to inc...
Page 41: ...nector Never pull or yank on the cord itself Connector Cord 4 5 5 Handpiece Light Setting You can switch the handpiece light ON OFF and change the brightness Page 1 SCALING SOFT AS USER A OFF L H Hand...
Page 42: ...g Clean The root canal length of the tooth in which the root is absorbed due to a lesion or which has incompletely developed root is indicated shorter and cannot be accurately measured Root Canal with...
Page 43: ...ival tissue In this case widen the opening at the top of the crown so that the file will not touch the metal prosthesis before taking a measurement Crown or metal prosthesis touching gingival tissue C...
Page 44: ...the components o in running water with a soft brush And then wipe o the water Cleaning Disinfection Packing Sterilization Be sure to follow the procedure below when performing autoclaving 2 Disinfect...
Page 45: ...nts Temperature 10 C to 70 C 14 F to 158 F Humidity 10 to 85 without condensatin Atmospheric Pressure 70kPa to 106kPa Avoid frequent or continuous exposure to direct sunlight If the equipment has not...
Page 46: ...Press the Reference button to make sure it changes from 1 to 2 to 3 etc Check that the probe cord can be properly plugged into its jack on the shoulder of the chair Check that the file holder s plug...
Page 47: ...rently making a measurement Reference numbers cannot be changed while making a measurement Switch may not be working In this case the switch may be defective Two or more bars on the meter may be blink...
Page 48: ...a metal prosthesis Touching a metal prosthesis with the file allows a flow of current to the gingival tissue or periodontal pocket and will cause the meter to jump to the Apex Is the tooth surface in...
Page 49: ...ealer or J MORITA OFFICE for inspection or repair Wire disconnection and circuit board fault Contact your local dealer or J MORITA OFFICE for inspection or repair 7 2 Error Display If the following er...
Page 50: ...on of the treatment instruments in the root canal is detected Mode of operation Continuous Power supply DC 5 V Power Rating 0 2 W Measurement Voltage AC 80 mV maximum Measurement Current 10 A maximum...
Page 51: ...U Built in Model may be repaired and serviced by The technicians of J MORITA s subsidiaries all over the world Technicians employed by authorized J MORITA dealers and specially trained by J MORITA In...
Page 52: ...of J MORITA CORP and J MORITATOKYO MFG CORP As the color of the product is printed it may differ from the actual color Some optional equipment is included Morita Global Site www morita com More Infos...