3
Section 1
Description
The Monmouth Powder Containment Booth provides local extraction and
containment of powders handled within the filter capture zone.
Refer to HSG258 for detailed guidance on application and suitable capture velocity.
The following capture velocities are achievable on this unit**:
Distance from centre of filter face Max Velocity (m/sec)
10mm
2.1
30mm
1.7
50mm
1.5
70mm
1.3
100mm
1.15
150mm
0.95
200mm
0.8
250mm
0.65
300mm
0.6
400mm
0.4
**When fitted with new filters
A 2-stage Filtration system is used, with 2 easy-change, front inlet grille mounted
course grade pre-filters to capture gross particulate and an H14 HEPA main filter to
provide higher-level containment.
PCB1800
External
Dimensions
1800mm Wide
750mmDeep
2400mm High
Internal
Dimensions
1700mmWide
600mmDeep
1900mm High