12
Set Normal Fan speed
– Opens a dialogue page that allows the user to set the fan
speed when the NORMAL button is pressed. Follow the on-screen prompts to set.
Set Fan speed 2
– Opens a dialogue page that allows the user to set the fan speed
when the 2
nd
fan speed button is pressed. Follow the on-screen prompts to set.
Be aware that an airflow lower than 0.3m/sec will trigger the low airflow alarm. If you
need to run the cabinet in this state then please recalibrate the alarm to trigger at a
point below your lowest setting.
Information
Pressing the
key on the main screen will display the information screen below
featuring key information about the unit.
Service information is reset by Monmouth personnel during a service visit. When a
service becomes due, a warning screen will be displayed momentarily on start-up
during the proceeding month.
Pressing the
key will display additional information of the filters fitted and UV tube
life. The
key will return to the previous screen.