3

Installation

Remove the mounting clips, if fitted, and install the unit into the panel from
the front.  From the rear of the unit, install the mounting clips on each side
and tighten the mounting screws against the rear of the panel.

WARNING:

Do not over tighten the mounting screws.

Power

Power to the unit is connected to the terminals under the section labeled

POWER 

on the rear panel. Be sure the power supplied matches the

specification indicated on the rear panel.  Refer to Figure 2 below.

Figure 2  ACT-1B Rear Panel

If the unit is 

ac powered

 (115 Vac or 230 Vac), connect the Live (Hot) wire

to the terminal marked 

L+

 and the Neutral (Return) wire to the terminal

marked 

N-

.  Connect the Ground (Earth) wire to the terminal marked 

GND

.

NOTE:

For full compliance with CE specifications, the Ground (

GND

)

connection must be made.

Analog Output Option (AO)

The analog output is 0 to 5 Vdc.

Connect the Positive side of the signal to the terminal marked 

OUT

,

and the Return side of the signal to the terminal marked 

COM

.

NOTE:

If your ACT-1B is equipped with either a current output or
an analog output, the full-scale output has been factory preset
to the speed range specified at the time of purchase. The output
is linear in 4096 discrete steps over its designated range.

Pulse Repeater Output Option (PO)

The Pulse Repeater output provides a conditioned TTL positive going
5 V pulse out for each pulse in.

Connect the Positive signal wire (+5 V pulse) to terminal marked 

+P

and the Return to the terminal marked 

COM

 in the rear panel section

labeled 

OUT

.

8

Summary of Contents for ACT-1B-10

Page 1: ...TRUMENT Instruction Manual 15 Columbia Drive Amherst NH 03031 2334 USA Phone 603 883 3390 Fax 603 886 3300 E mail techsup monarchinstrument com Website www monarchinstrument com Printed in the U S A M...

Page 2: ...all material respects when used in strict compliance with such documentation or manual Notwithstanding the foregoing BUYER acknowledges that since software is complex and therefore may have defects B...

Page 3: ...y that the product to which this declaration relates is in conformity with the following standards and therefore conforms with the requirements of Council Directive 89 336 EEC relating to electromagne...

Page 4: ...ed Sensors Optical Monarch ROS 5W Proximity Monarch P5 11 Magnetic Monarch M 190W or MT 190W Infrared Monarch IRS 5W Laser Monarch RLS 5W IO Option 4 to 20 mA see page 7 for maximum load calculation F...

Page 5: ...terminal connections on the rear panel of the instrument If specified at time of order placement the ACT 1B may be equipped with either an optional 4 to 20 mA current output IO or 0 to 5 Vdc analog ou...

Page 6: ...d N Connect the Ground Earth wire to the terminal marked GND NOTE For full compliance with CE specifications the Ground GND connection must be made Analog Output Option AO The analog output is 0 to 5...

Page 7: ...Contact the factory for increased sensitivity Connect the signal wire from three wire sensors or the positive side of two wire magnetic sensors to this terminal Typical input impedance is 10 Kohms COM...

Page 8: ...the rear panel CAUTION The IO or AO COM is NOT isolated from the other COM connections Current Output Option IO The current output is 4 to 20 mA This output is a current sink and requires a source vol...

Reviews: