19
GB
All rights reserved by MONACOR
®
INTERNATIONAL GmbH & Co. KG. No part of this instruction
manual may be reproduced in any form or by any means for any commercial use.
Voltage / Current / Resistance / Capacitance / Frequency
Function
Range
Resolution
Accuracy
DC voltage
Input impedance 7.8 MΩ
Overload protection 1000 V
¿
/~
400 mV
0.1 mV
±(0.5 % + 2 digits)
4 V
1 mV
40 V
10 mV
400 V
100 mV
1000 V
1 V
±(0.8 % + 2 digits)
AC voltage
Input impedance 7.8 MΩ
True RMS measurement, 50 – 60 Hz
Overload protection 1000 V
¿
/~
400 mV
0.1 mV
±(0.8 % + 3 digits)
4 V
1 mV
40 V
10 mV
400 V
100 mV
1000 V
1 V
±(1.2 % + 5 digits)
DC current
Fuse:
µ
A / mA range F 500 mAH / 1000 V
10 A range
F 10 AH / 1000 V
400
µ
A
0.1
µ
A
±(1.2 % + 3 digits)
4000
µ
A
1
µ
A
40 mA
10
µ
A
400 mA
100
µ
A
10 A
10 mA
±(2.5 % + 3 digits)
AC current
True RMS measurement, 50 – 60 Hz
Fuse:
µ
A / mA range F 500 mAH/1000 V
10 A range
F 10 AH/1000 V
400
µ
A
0.1
µ
A
±(1.5 % + 5 digits)
4000
µ
A
1
µ
A
40 mA
10
µ
A
400 mA
100
µ
A
10 A
10 mA
±(3.0 % + 5 digits)
Resistance
Overload protection 1000 V
¿
/~
400 Ω
0.1 Ω
±(0.8 % + 5 digits)
4 kΩ
1 Ω
±(0.8 % + 2 digits)
40 kΩ
10 Ω
400 kΩ
100 Ω
4 MΩ
1 kΩ
±(2.5 % + 8 digits)
40 MΩ
10 kΩ
Capacitance
Overload protection 1000 V
¿
/~
4 nF
1 pF
±(5.0 % + 20 digits)
40 nF
10 pF
±(5.0 % + 7 digits)
400 nF
0.1 nF
±(3.0 % + 5 digits)
4
µ
F
1 nF
40
µ
F
10 nF
400
µ
F
0.1
µ
F
4 mF
0.001 mF
±(10 % + 10 digits)
40 mF
10 mF
Frequency
≤ 1 MHz: sensitivity > 0,5 V
> 1 MHz: sensitivity > 3 V
Overload protection 1000 V
¿
/~
4 kHz
1 Hz
±(1.2 % + 3 digits)
40 kHz
10 Hz
400 kHz
100 Hz
10 MHz
1 kHz
±(1.5 % + 4 digits)