Assembly of the handpiece and the lever with electrode after cleaning
Figure 10
Figure 11
Figure 12
Figure 13
1. Place the lever with electrode on the handpiece and ensure that the lever is in the
vicinity of the handpiece (see
2. Ensure that the electrode is positioned parallel with the fixed electrode and is aligned
with the corresponding incision in the handpiece (see
).
3. Pull the lever with the electrode on the handpiece until both pins emit an audible
and
4. Press the lever twice and make sure that the electrodes move uniformly and are not
loose.
After cleaning, check the electrodes for signs of mechanical damage and wear.
Never use damaged parts!
After assembly of the sealing handle, perform a few trial sealing operations to
test functionality.
Plunger pins (on
both sides)
Correctly engaged
(on both sides)