background image

Power requirements

Europe:  

230 VAC ±10%, 50 Hz, 1250 W, 750 VA average, 
2500 VA maximum

North America:   115 VAC ±10%, 60 Hz, 1250 W, 750 VA average, 

2500 VA maximum

Japan: 

 100 VAC ±10%, 50/60 Hz, 1250 W, 750 VA average, 
2500 VA 

Facility receiving requirements

Only Molecular Devices personnel are authorized to open 
the crate(s). 

Equipment will arrive in 2 crates. Shipping crate dimensions 
and weights are shown below. Crates must be stored inside 
and protected from environmental extremes until equipment is 
removed. The instrument crate requires 10 feet of clearance on 
one side to allow removal of the instrument from the crate.

L

W

H

Weight

QPix 420 – depending on options it can have 2 or 3 crates

Instrument crate

104 cm

(41”)

1678 cm

(66”)

112 cm

(44”)

263 Kg

(579 lbs)

Accessory crate

94 cm

(37”)

117 cm

(46”)

97 cm

(38”)

162 Kg

(357 lbs)

Table/compressor 
crate

81 cm

(32”)

157 cm

(62”)

114 cm

(45”)

225 Kg

(496 lbs)

QPix 450 or 460 – 2 crates

Instrument crate

41 cm

(41”)

239 cm

(94”)

112 cm

(44”)

400 Kg

(879 lbs)

Accessory crate

127 cm

(50”)

218 cm

(86”)

112 cm

(44”)

350 Kg

(770 lbs)

Utility, Other Service, Safety Precautions and 
Hazard Identification sections

Compressed air

 

Compressed air is required for the picking movement of the 
picking pins. The optional oil-free compressed air unit draws air 
from the local environment and delivers it to the instrument through 
the regulator set to 100 psi (0.69 MPa).

Transport and storage

 

The robot must be stored and transported in temperatures within 
the range -25°C to +55°C. 

The machine should not be moved after installation. If relocation 
is necessary, standard lifting gear is adequate but should 
be undertaken only in the presence of a Molecular Devices 
approved engineer.

The robot should be moved into position using appropriate 
handling equipment such as forklift trucks or dolly trucks. The 
machine should be properly balanced on the forks prior to lifting.

Important note: Do not use any part of the exterior bodywork to 
lift the instrument as this may cause irreparable damage.

Front view of the QPix 450 Colony Picking System or QPix 460 Colony 
Picking System with dimensions. (1) 

Height of instrument with stacker 

cassettes: 122 cm (48.0 in.).

 (2)

 Height of instrument without stacker 

cassettes: 78 cm (30.7 in.). 

(3)

 Width of instrument: 217 cm (85.4 n.).

Side view of the QPix 450 Colony Picking System or QPix 460 Colony 
Picking System with dimensions. (1) 

Height of instrument with stacker 

cassettes: 122 cm (48.0 in.).

 (2)

 Height of instrument without stacker 

cassettes: 78 cm (30.7 in.). 

(3)

 Depth of instrument: 79 cm (31.1 in.).

Side view of the QPix 420 Colony Picking Systems with dimensions. 
(1) 

Height of instrument: 78 cm (30.7 in.).

 (2)

 Depth of instrument: 79 cm 

(31.1 in.).

Reviews: