ImageXpress Confocal HT.ai High-Content Imaging System User Guide
218
5084773 A
Configuring Parfocality After Changing Objectives
To configure parfocality, use the MetaXpress software to find valid focus values for each
objective, and then enter them into the MetaXpress Objective dialog in the Meta Imaging
Series Administrator Software. You can configure parfocality with or without the taskbar.
CAUTION! You must contact Molecular Devices Technical Support to schedule a Field
Service Engineer to install or remove a water immersion objective.
Configuring Parfocality With the Taskbar
If the IX Taskbar is not available, see
Configuring Parfocality Without the Taskbar on page 219
.
To configure parfocality:
1. In the IX Taskbar, click System Maintenance.
Note: If the Main Taskbar is not visible, press F4.
Figure 8-26: IXM Main Taskbar
2. Click Parfocality and XY Offsets and follow the prompts in the dialogs that display.
Figure 8-27: System Maintenance Taskbar
Summary of Contents for ImageXpress Confocal HT.ai
Page 1: ...5084773 A March 2021 ImageXpress Confocal HT ai High Content Imaging System User Guide...
Page 6: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 6 5084773 A...
Page 16: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 16 5084773 A...
Page 26: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 26 5084773 A...
Page 62: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 62 5084773 A...
Page 188: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 188 5084773 A...
Page 248: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 248 5084773 A...