Data Processing Algorithms
276
0112-0109 H
Each well-specific spatial uniformity correction factor is calculated by
dividing the mean fluorescence counts of all wells by the fluorescence
counts of each well (taken at Sample 1.) The table above also presents
the correction factor for wells A1–A9.
All samples taken from a particular well are multiplied by their well-
specific correction factor. For example, all samples from A1 are
multiplied by 1.12, A2 by 1.05,
etc
.
The results of applying the spatial uniformity correction factor are
presented in the table below. Note the decreased variability range of
wells A1–A9 in Sample 1 (8900–8976) as compared to the same data
prior to applying the correction algorithm (8000–10000).
If the spatial uniformity correction factor is applied to plates with empty
wells, non dye-loaded cells, or a panel of cells containing different dyes
and/or dye concentrations, the well-specific fluorescence counts will be
skewed by the correction factor. However, the EC
50
of the agonists
tested will not be affected.
Negative Control Correction
The negative control correction algorithm corrects for changes in
fluorescence that occur in all wells over the course of the experiment.
Causes for these changes in fluctuations in fluorescence include dye
leakage from cells, fluid addition artifacts, changes in illumination
power, dye photo-bleaching, and temperature drifts.
Sample Time
Well
A1
– Ctr
A2
– Ctrl
A3
– Ctrl
A4
Exp
A5
Exp
A6
Exp
A7
+ Ctrl
A8
+ Ctrl
A9
+ Ctrl
1
0
8960
8925
8930
8938
8924
8976
8900
8930
8961
2
5
9194
9030
8836
9047
8730
9180
8455
9118
9270
3
10
9408
8925
8648
9265
8730
9282
8633
9118
9682
4
20
9632
8820
8648
49050
48500
42840
50730
48880
55620
5
25
9632
9240
8648
40330
40740
35700
47170
47000
51500
6
30
9856
8925
8930
32700
33950
29580
44500
47940
51500
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...