3. Operating Instructions
Figure 3.41.
Smith chart of reactive 200 ohm load showing effect of reactive impedance
at higher frequencies. Cursor (red arrow) shows impedance value calculated at DC (0
MHz).
information will be taken from the time-windowed S
11
trace.
Figure 3.42 shows a normalized TDR trace (blue) with a short fault in a 50 ohm cable. The
normalized trace shows reduced aberrations and more accurate relative reflection coefficient and
impedance after the fault compared with the live TDR trace.
Figure 3.42.
Normalized trace (blue, bottom) showing short fault in a 50 ohm cable.
3.11.22. Vertical Units
The CT100B is able to change vertical units of the live TDR trace so that it is displayed in
units of reflection coefficient (millirho), impedance (ohms) or voltage standing wave ratio
(VSWR). To change vertical units, use the following procedure:
1. Select
MENU
→
Measurement
menu.
CT100B TDR Cable Analyzers Operator’s Manual
57