C.4 Camera settings
35
PULSED
MEAS,PULSED[,onoff]
Enable or disable “pulsed” measurement mode (see
TRIG
MEAS,TRIG[,onoff]
Enable or disable externally triggered mode, where a measurement
will only occur when the associated rear-panel SMA input is low
(see
4.5). The special argument keyword
SOFTWARE
will generate a
software trigger.
RATE
MEAS,RATE
Returns
the
current
rate
at
which
wavelength
measurements
are
being calculated, in measurements per second.
AVERAGE
MEAS,AVERAGE[,val]
Specify the wavelength measurement to average over
val
millise-
conds for improved measurement precision at the expense of measu-
rement rate.
If
val
is zero, no averaging is performed.
CLEAR
MEAS,CLEAR
Reset the measurement averaging and internal verification, for use
in combination with an external optical switch or shutter.
CORRECT
MEAS,CORRECT,val
Apply a correction to the device internal calibration using the currently-
supplied laser as an absolute reference.
The reference value
val
should be specified
in THz
and as many significant figures as practi-
cal. Perturbing the calibration in this way will improve the absolute
accuracy of the device around the reference wavelength, at the ex-
pense of the absolute accuracy far from the reference wavelength.
Specify
val
as the string “
RESET
” or “
FACTORY
” will revert the cali-
bration to the factory-provided values.
C.4
Camera settings
AUTO
CAM,AUTO[,value]
Enable or disable the auto-exposure algorithm to dynamically adjust
the exposure time to match the input (see
Summary of Contents for FZW600
Page 1: ...Fizeau Wavemeter FZW600 Version 0 1 2 Rev 2 hardware ...
Page 20: ...14 Chapter 3 User interface ...
Page 30: ...24 Chapter 4 Operation ...
Page 36: ...30 Appendix A Specifications ...
Page 44: ...38 Appendix C Command language ...
Page 49: ...E Dimensions 83 120 146 32 43 ...
Page 50: ...44 ...
Page 51: ......