background image

Health Check  

 

Press buttons 

MEM/SET

 and 

MODE

 together for 2 seconds to enter the countdown. Use the 

MODE

 button to toggle between 15 minutes, 30 minutes or 45 minutes or OFF. 

 

Press the 

MEM/SET

 button when the desired time is on screen to start counting down. When the 

countdown is active and running in the background the “

” sign will flash on screen.

 

 

When the countdown begins, the user can press 

MEM/SET

’ and ‘MODE’

 button together for 2 

seconds to display the remaining time, then press 

‘SCAN’

 button to return to waiting mode or press 

‘MEM/SET’

 button to cancel the countdown. 

 

When the countdown is zero. Health check 

“ ”

 symbol will display on LCD and the 

unit will ‘BEEP’ 

to 

remind the user. To stop 

the notification’s and return to the home menu press a

ny button. 

Memory Function

 

 

This unit can recall up to 9 prior measurements stored in memory. 

 

Turn on thermometer, press the 

MEM

 button once, then press it again to show the last 

measurement.   

 

The symbols 

for ‘EAR’ or ‘FOREHEAD’ 

will appear with each measurement stored in memory to 

indicate whether a person’s forehead

 or ear temperature was taken. 

 

Each press of the same button recalls a previous measurement.  

 

Troubleshooting 

Problem 

Checklist 

Countermeasure 

No response/automatic reset 
when pulling out plastic insulator 
tab.

 

Batteries are low or dead 
Batteries installed in wrong 
polarity?  
Poor battery contact?

 

Change to new batteries 
Remove batteries and reinsert 
batteries correctly

 

 

Measured temperature is lower 
than 89.6°F/32.0°F or higher than 
109°F/42.9°C

 

Follow steps in user manual for 
correct measurements.  
Check the device is in the correct 
operating mode.

 

 

These Error messages indicate a 
problem with the hardware.

 

Contact Mobi Technologies, Inc. 
via contactin information at the 
end of this manual.

 

Summary of Contents for FEVERTRACK

Page 1: ...ion Weight Forehead operating Condition Ear and Object operating Condition Atmosphere pressure Service life Service life of the battery DC 3V 2 x AAA 89 6 F 109 3 F 32 0 C 42 9 C Within range of 95 0 F 107 6 F 35 0 C 42 0 C Accuracy 0 4 F 0 2 C Outside of stated range Accuracy 0 5 F 0 3 C 41 0 F 139 8 F 5 0 C 59 9 C 2 0 F 1 0 C 150 0 L x40 5 W x 50 0 H mm 2mm 61 0g without batteries 59 0 F 104 0 F...

Page 2: ... not an indicator of battery power Battery indicator will display constantly when battery level is low 1 Infrared Sensor 2 LCD Display 3 Power Scan Button 4 Probe Cap 5 MEM SET Button 6 Temperature Mode Selection 7 Battery Cover Installing Replacing Battery Remove the battery cover by sliding it toward the bottom of the thermometer Insert or replace 2 x 1 5 AAA batteries into the battery compartme...

Page 3: ...st 30 minutes prior An unexpected fluctuation of ambient temperature is likely to lower measuring accuracy External factors may influence ear and forehead temperatures particularly when the individual has been o Lying on the ear to be measured or has it covered o Exposed to very hot or cold temperatures o Swimming or bathing causing the ear canal to be wet o Recently exercising bathing or other ph...

Page 4: ...Ensure no hair is obstructing the unit then press the SCAN button you will hear a short beep and the temperature will be displayed Note Results shown from 89 6 F to 109 4 F 32 0 C to 43 C HI is displayed for all results above 109 4 F 43 C and LO displayed for result below 89 6 F 32 0 C Object temperature measurement Remove the forehead cap Press the MODE button to switch to object mode will show u...

Page 5: ...s unit can recall up to 9 prior measurements stored in memory Turn on thermometer press the MEM button once then press it again to show the last measurement The symbols for EAR or FOREHEAD will appear with each measurement stored in memory to indicate whether a person s forehead or ear temperature was taken Each press of the same button recalls a previous measurement Troubleshooting Problem Checkl...

Page 6: ...easurements indoors Determine if you might have held the thermometer in your hand too long which may affect its accuracy Place the thermometer in the room where the measurement is due to be taken let it stabilize for 30 minutes before use Check to see if the Low Battery icon is showing on the LCD screen Immediately replace the batteries with new ones Product Safety Warnings WARNING WARNING WARNING...

Page 7: ...tmospheric pressure Operating Temperature Serial number Operating Humidity Recyclable Manufacturer This device compiles with Part 15 of FCC federal commuinications Commission Rules IP22 The first number 2 Protected against solid foreign objects to 12 5mm Φ and greater The second number Protected against vertically falling water drops when enclosure titled up to 15 ...

Page 8: ...mplete with its original parts accessories along with a copy of the original purchase gift or registry receipt to us This warranty does not cover damage caused by any tampering accident misuse abuse improper maintenance unauthorized modification or connection to an improper power supply There will be a charge for repair caused for such damage This warranty excludes all incidental or consequential ...