Mobi Technologies 70113 Instruction Manual Download Page 7

Compliance and EMC Declaration 

 

Compliance 

Equipment Classification 

Safety Standards: EN 60601-1: 2006+A1:2013, EN 60601-1-2:2007 

Type Of Protection 

Internally powered equipment (battery power) 

Degree of Protection 

Non-applied part 

Front Panel and Case Labelling 

EN ISO15223-1:2012 

Temperature 

EN ISO80601-2-56:2012 

Home Healthcare Environment 

EN 60601-1-11:2010 

 

Guidance and Manufacturers Declaration 

 Electromagnetic Emission 

The Non-Contact Infrared Body Thermometer is intended for use in the electromagnetic environment specified below. The 
user of the Non-Contact Infrared Body Thermometer should ensure that it is used in such an environment. 

Emission Test 

Compliance 

Electromagnetic Environment 

 Guidance 

RF emissions CISPR 11 

Group 1  

The Non-Contact Infrared Body Thermometer uses RF energy 
only for its internal function. Therefore, its RF emissions are 
very low and are not likely to cause any interference in 
nearby electronic equipment. 

RF Emission CISPR 11 

Class B 

The Non-Contact Infrared Body Thermometer is suitable for 
use in all establishments, other than domestic and those 
directly connected to the public low voltage power supply 
network that supplies building used for domestic purposes.  

Harmonic Emissions IEC 
61000-3-2 

Not Applicable 

Voltage 
Fluctuations/Flicker 
Emissions IEC 61000-3-3 

Not Applicable 

Guidance and Manufacturers Declaration 

 Electromagnetic Immunity 

The Non-contact Infrared Body Thermometer is intended for use in the electromagnetic environment specified below. The 
user of Non-Contact Infrared Body Thermometer should ensure that it is used in such an environment. 

Immunity Test 

IEC 60601 Test Level 

Compliance 
Level 

Electromagnetic Environment 

 Guidance 

Electrostatic discharge 
(ESD) IEC 61000-4-2 

±6 kV contact 
±8 kV air 

±6 kV contact 
±8 kV air 

Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic material, the relative humidity 
should be at least 30% 

Electrical fast 
transient/burst IEC 
61000-4-4 

±2 kV for power supply lines 
±1 kV for input/output lines 

Not applicable 

Mains power quality should be that of a 
typical commercial or hospital 
environment. 

Surge IEC 61000-4-5 

±1 kV line(s) to line(s) 
±2 kV line(s) to earth 

Not applicable 

Mains power quality should be that of a 
typical commercial or hospital 
environment.  

Voltage dips, short 
interruptions, and 
voltage variations on 
power supply input lines 
IEC 61000-4-11 

<5% U

T

 (>95% dip in U

T

) for 0.5 

cycle 
40% U

T

(60% dip in U

T

) for 5 cycles 

70% U

T

(30% dip in U

T

)for 25 cycles 

<5% U

T

(>95% dip in U

T

) for 5 sec 

Not applicable 

Mains power quality should be that of a 
typical commercial or hospital 
environment. If the user of the Non-
Contact Infrared Body Thermometer 
requires continued operation during 
power mains interruptions, it is 
recommended that the Non-Contact 
Infrared Body Thermometer be powered 
from an uninterruptible power supply or a 
battery. 

Power frequency 
(50Hz/60Hz) magnetic 
field IEC 61000-4-8  

3 A/m 

3 A/m 

Power frequency magnetic fields should 
be at levels characteristic of a typical 
location in a typical commercial or 
hospital environment. 

Note: U

T

 is the a.c. mains voltage prior to application of the test level.  

Summary of Contents for 70113

Page 1: ... Auto Power Off Time Measuring Time Measuring Distance Stored Memory Capacity Operating Condition C F Adjusted Mode body mode Direct mode surface mode Forehead Axillary Body mode 95 0 F 107 6 F 35 0 C 42 0 C Surface Mode 32 0 F 212 0 F 0 0 C 100 0 C Body Mode 93 2 F 109 2 F 34 0 C 42 9 C Surface Mode 32 0 F 212 0 F 0 0 C 100 0 C Body mode 0 4 F 0 2 C Surface mode 3 6 F 2 0 C 0 1 C 0 1 F Green Norm...

Page 2: ...the battery cover Insert two AAA batteries to match the and as illustrated in the battery compartment Close the battery cover and ensure it is locked in place before operation When replacing batteries always replace with new AAA batteries ensuring to again match polarities indicated in the battery compartment If the thermometer does not power on after replacing the batteries please remove the batt...

Page 3: ...ed ON device will beep and display sound icon on the screen Measurement Preparation For Best Results For more accurate measurements always store the device in same environment it is being used for at least 30 minutes prior An unexpected fluctuation of ambient temperature is likely to lower measuring accuracy External factors may influence forehead temperatures particularly when the individual has ...

Page 4: ... 2 F to 109 2 F 34 0 C to 42 9 C HI is displayed for all results above 109 2 F 42 9 C and LO displayed for result below 93 2 F 34 0 C Surface food temperature measurement Ensure the unit is in surface mode Press and release the MODE button to switch between modes surface will be visible on the display For optimal results point the unit towards the surface holding the probe 4 2 inches away from the...

Page 5: ...ll residual alcohol has evaporated before use Equipment Symbols Warning Compliance with WEEE Standard Non Sterile Packaging DO NO THROW AWAY intended for multiple use Refer to Operating Instructions Operating atmospheric pressure Operating Temperature Serial number Operating Humidity Recyclable Manufacturer This device compiles with Part 15 of FCC federal commuinications Commission Rules IP22 The ...

Page 6: ...ot use this thermometer if it has been malfunctioning or it has been broken or damaged This product is not waterproof Do not use this product under any wet conditions and do not immerse in any liquid or water This product should not be left alone with children or pets when not in use by an adult Do not store the thermometer in extreme temperatures below 4 F 20 C or over 131 F 55 C ...

Page 7: ...agnetic environment specified below The user of Non Contact Infrared Body Thermometer should ensure that it is used in such an environment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic ma...

Page 8: ...mplete with its original parts accessories along with a copy of the original purchase gift or registry receipt to us This warranty does not cover damage caused by any tampering accident misuse abuse improper maintenance unauthorized modification or connection to an improper power supply There will be a charge for repair caused for such damage This warranty excludes all incidental or consequential ...

Reviews: