MNR SV-2000 Operating Manual Download Page 2

SV-2000 OPERATING MANUAL 

 
 
Introduction 
 
The Solvent Vapor Generator is used to facilitate the evaporation of solvents in order to 
produce  an  airborne  vapor  stream.    It  can  be  used  to  produce  vapors  from  aqueous 
solutions  of  polar  species  (e.g.  formaldehyde).    Caution  must  be  exercised  when 
performing these functions. 
 
 
Vaporization of Pure Solvents 
 
Check  to  make  sure  that  you  are  operating  below  the  explosive  temperature  and 
concentration of the solvent being vaporized.  Also make sure that the solvent does not 
react  chemically  with  either  brass  or  copper.    Once  these  two  determinations  are  made, 
set up the vaporizer as follows: 
 

1.

 

Set vaporizer on the bench and connect a flowing air supply (e.g. MNR HCS-401) 
to the air INLET connection. 

 
2.

 

Establish the desired flow rate and relative humidity from the flowing air supply. 

 
3.

 

Turn on the vaporizer and set the heater setting to the desired level.  DO NOT 
TURN ON THE VAPORIZER UNLESS AIR IS FLOWING THROUGH THE 
SYSTEM.  The heater setting should be just enough to completely evaporate the 
liquid.  Too much heat could decompose the liquid. 

 
4.

 

Fill an appropriate sized syringe with the solvent of interest.  Connect the syringe 
to a length of PTFE tubing that terminates in a needle.   

 
5.

 

Pierce the septum on the INJECTION PORT with the needle and push until the 
needle stops (about 2 ¼ inches). 

 
6.

 

Place the solvent-filled syringe into a syringe pump and adjust the heater setting 
until a smooth outflow of vapor is established.  If the output from the vaporizer is 
irregular, the flow from the vaporizer may be passed through a mixing chamber 
prior to use to obtain a more uniform mixture. 

 

 
 
 
 
 
 
 

 