MMC D-2615-11 Operating And Maintenance Instructions Manual Download Page 16

SECTION VII 

 

 

7.0 

CALIBRATION PROCEDURE: 

 

7.0.1  The Oxygen Sensor and associated electronics have been accurately calibrated at factory. 

The oxygen cell used, and the electronic circuits are extremely stable and in general 

require calibration only when the 9-volt battery or the oxygen cell is replaced. 

 

7.0.2  A simple way to perform push button calibration has been incorporated in the design 

which permits the operator to quickly bring the system into calibration using the oxygen 

content of normal air (20.9%) ±0.2% as standard. 

 

7.1 

METHOD OF CALIBRATION: 

 

Calibration should be performed in a normal air environment and if possible, at room temperature 

(68 to 78°F). 

 

7.1.1  Turn power on by depressing the power “ON/OFF” push button. Wait approximately 10 

seconds for display to stabilize. Display should read 20.9% 

±

0.2. If proper reading is 

obtained there is no need to calibrate. If an incorrect reading is obtained, look at the upper 

left corner of the display. If “LO BAT” appears, indicating a low battery condition, 

replace the 9-volt battery contained in the hub following the procedure given in Section 

6. 

 

Note:

  When a 9-volt battery is replaced or temporarily removed, the memory circuits 

within the hub may lose their charge and cause the display to produce an 

abnormally low reading. If all the charge has leaked off, the reading will be zero. 

 

7.1.2  To perform the simple push button calibration depress the “Display Light” push button 

labeled “1” and while holding it in depress the power “ON/OFF” push button labeled “2”. 

Now release both. The display will jump to a low-value and quickly climb to a correct 

reading of 20.9% 

±

0.2. The order of button depression must be followed in the order 

described. 

 

7.1.3  No further adjustment is required. 

 

Note:

 If the oxygen cell has reached end of life, calibration will not be possible. A lower 

display reading of random value will be obtained each time calibration is 

attempted. Cell replacement must be performed to re-establish normal usage. See 

sell replacement in this manual. 

 

 

 

 

 

 

 

 

 

 

16 

 

Summary of Contents for D-2615-11

Page 1: ...SENSOR GAUGING TAPE CLOSED AND RESTRICTED APPLICATIONS MODEL D 2615 11 APPROVED BY BASEEFA AS INTRINSICALLY SAFE FOR USE IN CLASS I DIVISION I GROUPS C D HAZARDOUS ENVIRONMENTS MMC INTERNATIONAL CORP...

Page 2: ...S 7 4 2 RECOMMENDED SAFE USAGE CONDITIONS 7 4 3 FLUID LEVEL ULLAGE OF TANK OR OTHER VESSEL 8 V OPERATION 5 1 FAMILIARIZATION WITH OPERATING CONTROLS FEATURES LOCATED ON THE TAPE REEL HUB BEZEL NAMEPLA...

Page 3: ...CONTINUED FROM PREVIOUS PAGE VII CALIBRATION PROCEDURE 7 1 METHOD OF CALIBRATION 12 VIII FAULT FINDING 14 15 IX APPENDIX 16 X LIST OF APPLICABLE DRAWINGS 24...

Page 4: ...1000 Hours Depends on night light use Battery Drain Display Night Light off 1 0 MA IN AIR 0 7 ma with 8 Oxygen Night Light Battery Drain 9 0 ma use Storage Temperature 20 C to 40 C Tape Length Standar...

Page 5: ...ape Reel Position Lock Watertight Electronics Housing with Sealed Switches Stainless Steel Fasteners Nylon Coated Aluminum Frame for lightweight portability Low Battery Indicator Liquid Crystal Displa...

Page 6: ...replacing oxygen cells WARNING This equipment is solely intended for use in determining the percent by volume of the oxygen that may be present in a protective blanket of inert gas It is not claimed o...

Page 7: ...an 8 level 2 6 A single 9 volt battery contained within the hub assembly powers the oxygen sensor tape and sensor assembly Battery drain is extremely low approximately 0 75 milli amperes and if the un...

Page 8: ...may be used to give instant push button calibration of the system in air See section 7 0 3 5 When the oxygen sensor is powered in a normal ambient environment the reading of the digital will show an...

Page 9: ...flash slowly The horn is silent If the sensor is now lowered into a tank which contains an oxygen level below 18 but above 8 the horn will now emit a pulsating audible tone and the red lamp will then...

Page 10: ...CONDITIONS 4 2 1 The apparatus sensor should preferable be entered into a tank or vessel within an earthed sounding tube or pipe where such devices are provided and are normally used for temperature...

Page 11: ...liable means 4 3 2 With the fluid level known the oxygen sensor can be entered into the tank and unreeled to the mid point position of free space above and between the liquid level and tank top CAUTIO...

Page 12: ...NTIFICATION FUNCTION IDENTIFICATION FUNCTION DIGITAL DIS LAY PERCENT OXYGEN DISPLAYED LO EAT LOW EATI ERY INDICATOR L E D RSD RAPIDLY FT ASHES WHEN OXYGEN 0 LEVEL IS ABOVE 8 SLOWLY FLASHES AT LEVEL AB...

Page 13: ...turning 6 0 6 Never permit tape and sensor head to unwind freely control speed of descent by use of a restraining force on tape reel crank 6 1 BATTERY REPLACEMENT In Safe Area Only The battery should...

Page 14: ...he cell assembly Replace screen and retaining cap Then follow procedure given for cell calibration in Section 7 0 6 3 GAUGING TAPE REPLACEMENT In case of damage the gauging tape may be replaced by fol...

Page 15: ...aking sure that the tape numerals face in the same direction as the previous tape numerals faced 6 3 13 Use a silicone rubber sealant to re seal area around slotted tape entrance to reel hub Re wind t...

Page 16: ...ned look at the upper left corner of the display If LO BAT appears indicating a low battery condition replace the 9 volt battery contained in the hub following the procedure given in Section 6 Note Wh...

Page 17: ...is symptom is usually indicative of a faulty power On Off push button switch Check the switch with an Ohmmeter as explained in Problem No 1 above PROBLEM NO 3 When turned on in normal air unit does no...

Page 18: ...Hold the cell housing so that the V notch in the center ring of the connector is up See drawing A 2615 17 Connect the positive lead of a digital voltmeter capable of reading millivolts to pin from th...

Page 19: ...ay only make circuitry repair Unauthorized repairs will void any guarantee or warranty given elsewhere in this manual In addition to the above statement it is equally if not more important to understa...

Page 20: ...APPENDIX PAGE NUMBER A MMC Warranty Statement 21 B Oxygen Sensor Gauging Tape Warranty Addendum 22 C Glossary of Approvals 23 D BASEEFA Certificate of Conformity Ex93C2037X 25 28 20...

Page 21: ...upon discovery of the defects B Upon receipt of written authorization from the seller the equipment is to be returned as directed transportation prepaid by the buyer C Buyer is to disclose the use of...

Page 22: ...reements with the above mentioned approval agencies and yet accommodate our customers as much as possible we have developed four modules which can be purchased as units installed by the owners of the...

Page 23: ...SPECTION BODIES APPORVALS ABS American Bureau of Shipping CCS China Classification Society DNV Det Norske Veritas Norway LLOYD S Lloyd s of London U K BV Bureau of Veritas GAUGING TAPE AND OTHER MEASU...

Page 24: ...2615 17 Oxygen Sensor Cell Outline Drawing B 2615 14 Tape Reel Hub Electronics Module Assembly S 2615 OCX Flexi Dip Portable Tape Assembly Oxygen Sensor Closed Gauging DB 2615 1FRX Flexi Dip Portable...

Page 25: ...25...

Page 26: ...26...

Page 27: ...27...

Page 28: ...28...

Page 29: ......

Page 30: ......

Page 31: ......

Page 32: ......

Reviews: