4. PARAMETERS AND CALIBRATION
4.1 Entering setting mode
To set the parameters, turn on the weighing scales and hold the
TARE
key during self-
checking, until the display show “F1 CAL”.
4.2 Calibration
After entering settings mode, and setting the parameter. The display will show "F1 CAL".
Press the
ZERO
key, display shows “UNLD”, empty the platter, after the “STABLE” indicator
turns on, press the
ZERO
key again. The display will show the calibrate weight. Press the
TARE
key to select the appropriate calibrate weight. After selection, put the calibrate weight
on. When the
“STABLE” indicator turns on, press the
ZERO
key to confirm. The display will
shows “pass”. The calibration will be complete.
4.3 Setting resolution
After entering setting mode, the display will show
“F1 CAL”. Press the
TARE
key, until the
display shows “F2 RES”, press the
ZERO
key to enter, and press the
TARE
key to select
the resolution.
4.4 Changing capacity
After entering the settings mode, the display will show
“F1 CAL”. Press the
TARE
key until
the
display shows “F3 CAP”. Press the
ZERO
key to enter, then the
TARE
key to select
capacity. Press
ZERO
key to confirm.
4.5 Checking internal counts
After entering the settings mode, the display will show
“F1 CAL”. Press the
TARE
key until
the
display shows “F4 INP”. Press the
ZERO
key to enter, display will show internal counts.
Press the
TARE
key to return settings.
4.6 Automatic sleep-mode
In weighing mode, if the scale is unmoved, and there is no change of weight, the scale will
enter sleep mode after a minute. Increasing the weight by 4d will re-wake the scale for use.