2
Initial Setup Procedures
2-13
Series 370 Stabil-Ion Instruction Manual - 370119
2.1.5.3 Selecting the Byte Format for RS-485 Communications
2.1.5.3.4 Baud Rate
Baud rate for the RS-485 communications is determined by the settings of switches S2.6, S2.7 and
S2.8. See Figure 2-15. Table 2-11, below, defines the baud rate base on the switch settings. The
factory default baud rate setting is 9600.
2.1.5.3.5 Character Framing for the RS-485 Computer Interface
Character framing for the RS-485 computer interface is determined by setting switches S2.3, S2.4,
and S2.5. See Figure 2-15. Table 2-12 below defines the character bits, parity, and stop bits based
on the switch settings. The factory default setting is S2.3 ON, S2.4
OFF
, and S2.5
OFF
- Character bits
set to 8, Parity None, and Stops bits at 1.
* factory setting
Table 2-11
Baud Rate Switch Settings
S2.6
S2.7
S2.8
Baud Rate
ON*
ON*
OFF*
9600*
ON
OFF
ON
4800
ON
OFF
OFF
2400
OFF
ON
ON
1200
OFF
ON
OFF
600
OFF
OFF
ON
300
OFF
OFF
OFF
150
Table 2-12
Character Framing Switch Settings
S2.3
S2.4
S2.5
Character Bits
Parity
Stop Bits
ON
ON
ON
8
None
2
ON
ON
OFF
8
Even
1
ON
OFF
ON
8
Odd
1
ON
OFF
OFF
8*
None*
1*
OFF
ON
ON
7
Even
1
OFF
ON
OFF
7
Odd
1
OFF
OFF
ON
7
Even
2
OFF
OFF
OFF
7
Odd
2
Summary of Contents for Stabil-Ion 370 Series
Page 2: ......
Page 8: ...Table of Contents iv Series 370 Stabil Ion Instruction Manual 370119...
Page 36: ...2 Initial Setup Procedures 2 16 Series 370 Stabil Ion Instruction Manual 370119...
Page 46: ...3 Installation 3 10 Series 370 Stabil Ion Instruction Manual 370119...
Page 58: ...3 Installation 3 22 Series 370 Stabil Ion Instruction Manual 370119...
Page 121: ......