InSight

®

 X3+

 Systems Pre-installation Guide 

InSight

®

 X3+™ and X3+ A™ laser systems

 

Spectra-Physics Lasers 

1565 Barber Lane 

Milpitas, CA 95035 

408-980-4300 

[email protected] 

[email protected] 

Summary of Contents for Spectra-Physics InSight X3+

Page 1: ...KS Spectra Physics www spectra physics com InSight X3 Systems Pre installation Guide InSight X3 and X3 A laser systems Spectra Physics Lasers 1565 Barber Lane Milpitas CA 95035 408 980 4300 sales spec...

Page 2: ...thin one week of your delivery date you will receive a call from one of our factory trained Customer Service Engineers Some preliminary planning is essential to avoid unnecessary delays during install...

Page 3: ...end as much time as possible reviewing the system components before your Spectra Physics Customer Service Engineer arrives for the installation and training Refer to Chapter 4 of the User s Manual for...

Page 4: ...m operating room temperatures are 20 25 C Room temperature should ideally be 22 C and should not fluctuate more than 1 C during any two hour period See specifications in Table 1 1 and table 1 3 in the...

Page 5: ...00 240VAC 50 60 Hz Current 10A 6A max Heat Load is 1 44kW max at 20C Chiller SSC ThermoRack 401 Voltage 120 240VAC 50 60 Hz Current 9A max Heat Load is 2 16kW max at 20C 2 4 Computer Requirements Two...

Page 6: ...ngineer s arrival unless specifically told to do so by the factory Two or more people will be required to lift some equipment _ Compare the packing list with your quotation Call your Spectra Physics o...

Page 7: ...the pink tracer turns clear This solution is a complete inhibitor that protects ferrous metals and copper alloys from corrosion It is nitrite free and minimizes the challenge of bacteria control depe...