Temperature, operating 

20

 

ºC to 25

 

ºC 

Relative humidity, operating 

Maximum 75% noncondensing, up to 25 ºC 

Temperature, storage 

20

 

ºC to 25

 

ºC 

Relative humidity, storage 

< 65% for 15ºC to 35 ºC 

Cooled water temperature in closed-loop 

chiller 

21ºC typical

1

 

1. Avoid obstructing the air exhaust grills which will result in the recirculation of hot exhaust air. Cooling air 
enters through the front panel and exits through the rear fan apertures. 

1.3 Utility Requirements 

Chiller and Cooling Liquid 

The InSight

®

 X3

TM

 has a closed-loop liquid cooling system and requires no utility water services.  Use 

ONLY  full-strength Nalco Coolant (P/N 1607-0546) that contains corrosion and bacterial/algal 
growth inhibitors. It is included in the InSight X3 crate. 

Never use deionized water, it can cause 

damage to the laser

.  

Fill  the chiller reservoir following the instructions in the chiller manufacturer's 

user manual. 

The chiller should be placed below the laser in a position where if a leak were to develop it would 
not damage the laser or other components on the optical table. The chiller must always be on when 
the power supply is on, even if the laser is not running. 

Electrical Power Requirements 

It is recommended to connect the laser and chiller on separate AC power circuits. 

 

Unit 

Maximum Power Requirement 

IPS300 Power Supply 

100 – 120 V~/ 10 A max. 

200 – 240 V~/ 6 A max. 

50 / 60 Hz 

ThermoRack 401 Chiller 

100 – 240 V~ / 9 A max. 

50/60 Hz 

 

Room Temperature (ºC) 

Typical Heat Generation 

Maximum Heat Generation 

20

 

ºC 

775W 

1050W 

25

 

ºC 

1100W 

1350W 

Summary of Contents for Spectra-Physics InSight X3

Page 1: ...MKS INSTRUMENTS INC SPECTRA PHYSICS LASERS 3635 PETERSON WAY SANTA CLARA CA 95054 2809 USA P 1 408 980 4300 WWW MKSINST COM WWW SPECTRA PHYSICS COM Pre Installation Guide InSight X3TM...

Page 2: ...eady to ship you will receive a call or email from one of our customer service engineers to plan out the delivery and installation of the system by one of Field Service Engineers Please follow the ste...

Page 3: ...s 22 C with fluctuations less than 1 C during any two hour period F In most applications vibration isolation is required and structural integrity of the floor will impact system performance G The lase...

Page 4: ...ion and bacterial algal growth inhibitors It is included in the InSight X3 crate Never use deionized water it can cause damage to the laser Fill the chiller reservoir following the instructions in the...

Page 5: ...ements Intel or AMD 32 or 64 bit single or multi core processor with 1 GHz clock speed 1 GB RAM 32MB available disk space for installation a CD ROM or DVD ROM drive a mouse or other Windows compatible...

Page 6: ...ase file a claim with the carrier and notify Spectra Physics The InSight X3 laser is designed to be installed on site by a factory trained installation engineer Do not open the laser packaging prior t...

Page 7: ...el In particular please refer to Chapter 4 of the user manual for more information on installation 3 3 Laser Safety Considerations In addition to reviewing the sections in the manual regarding laser s...

Page 8: ...ing solution P N 1607 0546 is a premixed liquid corrosion inhibitor bacterial growth inhibitor and cooling liquid designed for use in the closed closed loop cooling system on the InSight X3 Laser Syst...

Page 9: ...r lifetime 4 4 General Procedures The Chiller output power and general performance of the laser system should be checked regularly If any of the output characteristics have changed please call Spectra...

Reviews: