4
Temperature, operating
20
ºC to 25
ºC
Relative humidity, operating
Maximum 75% noncondensing, up to 25 ºC
Temperature, storage
20
ºC to 25
ºC
Relative humidity, storage
< 65% for 15ºC to 35 ºC
Cooled water temperature in closed-loop
chiller
21ºC typical
1
1. Avoid obstructing the air exhaust grills which will result in the recirculation of hot exhaust air. Cooling air
enters through the front panel and exits through the rear fan apertures.
1.3 Utility Requirements
Chiller and Cooling Liquid
The InSight
®
X3
TM
has a closed-loop liquid cooling system and requires no utility water services. Use
ONLY full-strength Nalco Coolant (P/N 1607-0546) that contains corrosion and bacterial/algal
growth inhibitors. It is included in the InSight X3 crate.
Never use deionized water, it can cause
damage to the laser
.
Fill the chiller reservoir following the instructions in the chiller manufacturer's
user manual.
The chiller should be placed below the laser in a position where if a leak were to develop it would
not damage the laser or other components on the optical table. The chiller must always be on when
the power supply is on, even if the laser is not running.
Electrical Power Requirements
It is recommended to connect the laser and chiller on separate AC power circuits.
Unit
Maximum Power Requirement
IPS300 Power Supply
100 – 120 V~/ 10 A max.
200 – 240 V~/ 6 A max.
50 / 60 Hz
ThermoRack 401 Chiller
100 – 240 V~ / 9 A max.
50/60 Hz
Room Temperature (ºC)
Typical Heat Generation
Maximum Heat Generation
20
ºC
775W
1050W
25
ºC
1100W
1350W