Pressure Level Comparators
Chapter Four: Theory of Operation
30
Pressure Level Comparators
PDR-C-2C only
The channel 1 pressure signal is applied to two comparators.
One of these is set to trip at 90% of full scale pressure while the other is set for 100%. The
comparator outputs are connected to logic whose output goes to the front panel Channel
Selector/Function Switch.
Channel Selector / Function Switch
PDR-C-2C only
This switch is used to determine which channel’s output is displayed by the front panel DVM by
choosing which logic level or input is applied to the channel select relay (K1). If the AUTO switch
position is chosen then the logic output from the pressure level comparators is applied to the relay
switching transistor. Front panel LEDs are used to indicate which channel is being displayed.
When the Channel Selector/Function Switch is set to the REMOTE position the user can remotely
select which channel is being displayed by inserting the appropriate logic level on pin C of the rear
panel Interface connector.