background image

12

Ionization Gauge Instruction Manual - 274026 - Rev. B

2.3

Electrical Connections

Do not use gauge cables with exposed conductors such as alligator clips. 
All gauge pins must be covered by connectors or by pin covers (insulation). 
Gauge cables should be firmly clamped to the vacuum station to provide 
strain relief and ensure that minimal strain is transmitted to the gauge pins 
if there is motion between the gauge and the gauge controller.

If a resistance heated (I

2

R) degassable nude gauge is being used with an 

electron bombardment (EB) controller, be sure that the unused grid pin on 
the gauge is covered (insulated).

Figure 2-1 lists the two types of electrical connectors available on the nude 
gauges (near the botttom of the chart).

Figure 2-2 illustrates the pin connections for Series 274 B-A gauges.

Figure 4-1 illustrates the nude gauges with and without the strain relief for 
the electrical connector.

2.4

Grounding Requirements

All conductors in, on, or around the vacuum system that are exposed to 
potential high voltage electrical discharges must either be shielded at all 
times to protect personnel or must be connected to earth ground at all 
times.

Connect a heavy duty #12 AWG or larger ground wire from the ground lugs 
on the backs of the Power Supply and the Control Unit to your facility 
grounding electrode system. This will provide an earth ground for the Power 
Supply and for the Control Unit in the event either the interconnect cable 
or power cable are not in place. Do not connect the ground lug to the 
vacuum system or other component. Connect it directly to the facility 
grounding system such as a grounded outlet box or a grounded copper 
water supply pipe. Do not rely on small metal water lines to ground a 
component. Someone may replace the metal tubing with plastic tubing thus 
unwittingly causing a potentially dangerous situation.

Provide a connection to ground for other instruments with electrodes in the 
vacuum system possibly exposed to high voltage electrical discharges.

Provide a connection to ground for each ungrounded metal component in, 
on or around the vacuum system, including the gauge envelopes, which 
personnel may touch and which can potentially be exposed to high voltage 
electrical discharges within the vacuum system. For example, a metal bell 
jar resting on an organic O-ring must be connected to ground if an 
ionization gauge is to be used or if other high voltage sources are present in 
the vacuum system.

Summary of Contents for Granville-Phillips 274 Series

Page 1: ...Granville Phillips Series 274 Bayard Alpert Type Ionization Gauges Instruction Manual part number 274026 Revision B August 2014 Series 274 Instruction Manual...

Page 2: ...Series 274 Tubulated and Nude Bayard Alpert type Ionization Gauges See Figure 2 1 on page 13 Theory of Operation on page 15 and Table 4 4 on page 28...

Page 3: ...respective owners Granville Phillips Series 274 Bayard Alpert Type Ionization Gauges Series 274 Instruction Manual For Customer Service or Technical Support 24 hours per day 7 days per week every day...

Page 4: ...log numbers for Granville Phillips Tubulated Glass and Nude Ionization Gauges Detailed information regarding the catalog numbers listed below is provided in Figure 2 1 on page 13 and Table 4 4 on page...

Page 5: ...Inspection of Products 11 2 2 Vacuum System Connections 11 2 3 Electrical Connections 12 2 4 Grounding Requirements 12 Chapter 3 Operation 15 3 1 Theory of Operation 15 3 2 Filament Emission 17 3 3 P...

Page 6: ...6 Ionization Gauge Instruction Manual 274026 Rev B...

Page 7: ...s instruction manual explain the specific hazard that you must prevent or unsafe practice that you must avoid the potential result of your failure to prevent the specified hazard or avoid the unsafe p...

Page 8: ...rt you to hazards or unsafe practices that could result in minor personal injury or property damage Each caution statement explains what you must do to prevent or avoid the potential result of the spe...

Page 9: ...sure at a level considerably below that pressure which the system can safely withstand Do NOT use quick connects or other friction type connections where positive pressure such as backfilling operatio...

Page 10: ...hock could then occur because of the high voltage between these exposed conductors and ground 1 8 Operation It is the installer s responsibility to ensure that the automatic signals provided by the pr...

Page 11: ...gauge should be mounted as close as possible to the section of the vacuum chamber where pressure measurement is important Valves or other restriction between the gauge and the area where pressure meas...

Page 12: ...to earth ground at all times Connect a heavy duty 12 AWG or larger ground wire from the ground lugs on the backs of the Power Supply and the Control Unit to your facility grounding electrode system T...

Page 13: ...Gauges are electrically equivalent to other manufacturers but the pin geometry may be different Granville Phillips Nude Gauges catalog numbers 274022 274023 274028 and 274050 do not have a strain rel...

Page 14: ...Chapter 2 14 Ionization Gauge Instruction Manual 274026 Rev B Figure 2 2 Series 274 Bayard Alpert Ion Gauge Connectors...

Page 15: ...e is based on the ionization of the gas molecules by a constant flow of electrons The negative electrons are emitted at a well controlled selectable rate from a heated filament cathode and are acceler...

Page 16: ...cs The recommended operating voltage potentials are Collector 0 V Grid 180 Vdc Filament 30 Vdc See the Operating Characteristics Charts for variations of these parameters Tungsten Helical Filament cat...

Page 17: ...ffs on emission current are that high emission current gives better readout stability and sensitivity but more pumping if the gauge is clean and more outgassing if the gauge is contaminated Likewise l...

Page 18: ...ose Figure 3 3 Operating Characteristics for Series 274 Bayard Alpert Gauges 3 4 X ray Limit The low end of the operating range of a B A gauge is determined by the X ray limit of this type of gauge Th...

Page 19: ...deposits can generally be removed by degassing the electrodes 3 5 Degas Procedure The deposition of elements or compounds on exposed gauge surfaces can result from some processes such as sputtering o...

Page 20: ...degassable nude gauges 274022 274023 274041 274042 274050 274057 and 274058 See the NUDE gauges listed in Figure 2 1 Tubulated glass gauges and nude gauges 274028 and 274043 can be degassed using eit...

Page 21: ...indicator for pressures below 1 x 10 3 Torr More detailed information regarding the theory of operation for B A gauges is presented in the Operation Chapter of this Instruction Manual The low end of t...

Page 22: ...sign permit operation to various orders of vacuum The nude type ionization gauges 274022 274023 274041 274042 274050 274057 and 274058 are suitable to 2 x 10 11 Torr Table 4 1 Operating Ranges of Baya...

Page 23: ...ament to the grid and P is the pressure The glass envelope gauge tube and the I2R degassable nude gauge have K 10 Torr for nitrogen or air and the electron bombardment degassable nude gauge has K 25 T...

Page 24: ...6 Rev B To convert the readout to a different gas the equation must also include KgasX tube 2 Most ion gauge controllers are designed for some calibrated set point of emission i_ The emission that cor...

Page 25: ...Literature from NASA Technical Note TND5285 by Robert L Summers Lewis Research Center National Aeronautics and Space Administration Refer to this reference for further definition of these average valu...

Page 26: ...Ionization Gauges UHV Nude B A Gauge EB Degas Nude B A Gauge I2R or EB Degas Tubulated B A Gauge I2R or EB Degas Nude ion gauge without cable connector strain relief Catalog numbers 274022 274023 and...

Page 27: ...l thoria coated Iridium on replaceable assembly Thoria coated iridium on replaceable assembly Grid Refractory metals Refractory metals Refractory metals Overall Length 6 inch 152mm 4 125 inch 105mm 4...

Page 28: ...ardous materials document is required When returning a products to Granville Phillips be sure to package the products to prevent shipping damage Circuit boards and modules separated from the gauge ass...

Page 29: ...Instruction Manual part number 274026 Revision B August 2014...

Page 30: ...port 24 hours per day 7 days per week every day of the year including holidays Phone 1 800 227 8766 or 1 303 652 4691 MKS Granville Phillips Division 6450 Dry Creek Parkway Longmont CO 80503 USA Phone...

Reviews: