Chapter Two: Installation
Setup
9
Note
1.
The grounds of any external power supply AND readout should be
the same as the sensor ground (input ports) to avoid any possible
ground loops which can cause noise in the system.
2.
Power requirements are less than 25 mA at either ±12 or ±15 Volt
supplies. Actually, the supply can be any voltage from 11 to 18
Volts. The output may be loaded with 10K ohms or greater.
Figure 2: Electrical Connections
Summary of Contents for Baratron 223B
Page 3: ...107876 P1 Rev C 8 96 MKS Baratron Type 223B Pressure Transducer ...
Page 6: ...Table of Contents iv ...
Page 8: ...List of Figures List of Tables vi ...
Page 18: ...Setup Chapter Two Installation 10 This page intentionally left blank ...
Page 20: ...Circuit Description Chapter Three Overview 12 Figure 3 Schematic of a Type 223 Transducer ...
Page 22: ...How To Adjust the Zero Pot Chapter Four Operation 14 This page intentionally left blank ...
Page 24: ...General Information Chapter Five Repair 16 This page intentionally left blank ...
Page 30: ...Model Code Appendix B Model Code Explanation 22 This page intentionally left blank ...